We have characterized the performance of a commercial two-site immunoradiometric assay for manual in vitro diagnostic measurement of plasma corticotropin from Scantibodies Laboratory. We compared the results with those of a similar commonly used assay from Nichols Institute Diagnostics that has recently been withdrawn from production. The lower detection limit, range of the standard curve, cross-reactivity, and intra-assay and inter-assay imprecision of the two assays are very similar. Measurement of clinical samples and a series of samples from an experimental subject demonstrate high correlations between the two assays. These factors, together with recent clearance by the United States Food and Drug Administration for manual in vitro diagnostic measurement, make the Scantibodies corticotropin immunoradiometric assay an appropriate replacement for the Nichols assay.
机译:我们已经表征了从Scantibodies实验室对血浆促肾上腺皮质激素进行手动体外诊断测量的商业两点免疫放射分析的性能。我们将结果与最近从生产中撤出的Nichols Institute Diagnostics的类似常用测定法进行了比较。两种测定的下限,标准曲线范围,交叉反应性以及批内和批间不准确性非常相似。临床样品和来自实验对象的一系列样品的测量证明了两种测定之间的高度相关性。这些因素以及最近美国食品和药物管理局对手动体外诊断测量的批准,使Scantibodies促肾上腺皮质激素免疫放射分析成为Nichols分析的合适替代品。