A capacitively coupled contactless conductivity detector ((CD)-D-4) was used for the determination of three groups of ionizable species in reversed-phase HPLC with isocratic elution. These were non-steroidal anti-inflammatory drugs which are arylpropionic acid derivatives and represent anionic analytes, beta-blockers which are amines and therefore cationic, and zwitterionic amino acids. Optimization of the eluents led to detection limits in the order of 1 mu M for all species. The precision in peak areas was typically between 0.2 and 2. 1% and calibration curves were linear up to 500 mu M. The determination of ibuprofen, acetylsalicylic acid and atenolol in real samples was also demonstrated by direct injection of dissolved pharmaceutical formulations into the HPLC-(CD)-D-4 system. (c) 2005 Elsevier B.V. All rights reserved.
机译:使用电容耦合的非接触电导检测器((CD)-D-4)在反相HPLC中通过等度洗脱测定三组可电离物质。这些是非甾体抗炎药,是芳基丙酸衍生物,代表阴离子分析物;β-阻滞剂是胺,因此是阳离子氨基酸,两性离子氨基酸。洗脱液的优化导致所有物种的检出限约为1μM。峰面积的精确度通常在0.2到2之间。1%且校准曲线在500μM时呈线性。通过直接将溶解的药物制剂注入HPLC也可测定实际样品中的布洛芬,乙酰水杨酸和阿替洛尔-(CD)-D-4系统。 (c)2005 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号