The performance of an ion chromatographic method for measuring citrate and phosphate in pharmaceutical solutions is evaluated. Performance characteristics examined include accuracy, precision, specificity, response linearity, robustness, and the ability to meet system suitability criteria. In general, the method is found to be robust within reasonable deviations from its specified operating conditions. Analytical accuracy is typically 100 ± 3%, and short-term precision is not more than 1.5% relative standard deviation. The instrument response is linear over a range of 50% to 150% of the standard preparation target concentrations (12 mg/L for phosphate and 20 mg/L for citrate), and the results obtained using a single-point standard versus a calibration curve are essentially equivalent. A small analytical bias is observed and ascribed to the relative purity of the differing salts, used as raw materials in tested finished products and as reference standards in the analytical method. The assay is specific in that no phosphate or citrate peaks are observed in a variety of method-related solutions and matrix blanks (with and without autoclaving). The assay with manual preparation of the eluents is sensitive to the composition of the eluent in the sense that the eluent must be effectively degassed and protected from CO2 ingress during use. In order for the assay to perform effectively, extensive system equilibration and conditioning is required. However, a properly conditioned and equilibrated system can be used to test a number of samples via chromatographic runs that include many (> 50) injections.
机译:评估了用于测定药物溶液中柠檬酸根和磷酸根的离子色谱方法的性能。检查的性能特征包括准确性,精确度,特异性,响应线性,鲁棒性以及满足系统适用性标准的能力。通常,发现该方法在与其指定的操作条件有合理偏差的范围内是可靠的。分析精度通常为100±3%,相对于标准偏差的短期精度不超过1.5%。仪器响应在标准制备目标浓度的50%至150%范围内呈线性关系(磷酸盐为12 mg / L,柠檬酸盐为20 mg / L),并且使用单点标准液与校准曲线获得的结果基本相同。观察到较小的分析偏差,并将其归因于不同盐的相对纯度,这些盐用作测试的最终产品的原材料,并用作分析方法中的参考标准。该测定具有特异性,因为在各种方法相关的溶液和基质空白(有或没有高压灭菌器)中均未观察到磷酸盐或柠檬酸盐峰。手动制备洗脱液的方法对洗脱液的成分很敏感,因为在使用过程中必须对洗脱液进行有效脱气并保护其免受二氧化碳的侵入。为了使测定有效地进行,需要广泛的系统平衡和调节。但是,可以使用经过适当调节和平衡的系统通过包括许多(> 50)进样的色谱运行来测试许多样品。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号