Precision of biometry, keratometry, and refractive measurements with a partial coherence interferometry-keratometry device.

机译:使用部分相干干涉法-角膜测量仪的生物测定,角膜测定法和屈光测量的精度。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To evaluate the precision of the axial length (AL), keratometry (K), anterior chamber depth (ACD), astigmatism, and minus astigmatic cylinder axis measurements by a partial coherence interferometry (PCI)-keratometry device. SETTING: Private practice, Lynwood, California. METHODS: This prospective comparative observational study analyzed measurements in the second eye to have cataract surgery. Before surgery in the first eye, AL, K, ACD, astigmatism, and cylinder axis in both eyes were measured with an IOLMaster PCI device. The measurements were repeated approximately 1 month later, before second eye-surgery. The 2 sets of measurements were compared. RESULTS: The study evaluated 121 eyes of 121 patients. The interclass correlation coefficient (ICC) for AL was 0.999 in all 3 signal-to-noise ratio (SNR) categories; the highest difference range was with an SNR below 100. Astigmatism, K, and cylinder axis had a high correlation in flat corneas (K reading <42.0 diopters [D]) (ICC = 0.994, 0.978, and 0.918, respectively) and a poorer correlation with K readings between 42.0 D and 44.0 D (ICC = 0.905, 0.774, and 0.456, respectively) and K readings above 44.0 D (ICC = 0.988, 0.729 and 0.446, respectively). CONCLUSIONS: The precision of the PCI measurements was extremely high for AL with low fluctuations (95% limits of agreement [LoA], 0.06 mm) and was relatively high for K readings with higher fluctuations (95% LoA, 0.55 D) and for ACD (95% LoA, 0.2 mm). The precision of astigmatism and cylinder axis was high in flat corneas and relatively low in steeper corneas.
机译:目的:通过部分相干干涉术(PCI)-角膜曲率测定仪评估轴向长度(AL),角膜曲率(K),前房深度(ACD),像散和负像散圆柱轴测量的精度。地点:加利福尼亚州林伍德市的私人诊所。方法:这项前瞻性比较观察性研究分析了第二只接受白内障手术的眼睛的测量结果。在第一只眼进行手术之前,使用IOLMaster PCI设备测量两只眼睛的AL,K,ACD,散光和圆柱轴。在第二次眼科手术之前大约1个月后重复测量。比较两组测量值。结果:该研究评估了121例患者的121眼。在所有3种信噪比(SNR)类别中,AL的类间相关系数(ICC)为0.999;最高差异范围是SNR低于100。散光,K和柱面轴在平坦角膜中具有高度相关性(K读数<42.0屈光度[D])(ICC分别为0.994、0.978和0.918),并且差得多与K读数在42.0 D和44.0 D之间(ICC分别为0.905、0.774和0.456)和K读数在44.0 D以上(ICC分别为0.988、0.729和0.446)相关。结论:PCI测量的精度对于波动较小的AL(95%的协议限值[LoA],0.06 mm)极高,而对于波动较大的K读数(95%LoA,0.55 D)和ACD而言,PCI测量的精度相对较高(95%LoA,0.2毫米)。散光和柱面轴的精度在平角膜中较高,而在较陡的角膜中相对较低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号