To determine the appropriate correction of predicted intraocular lens (IOL) power for sulcus fixation and to prospectively assess the validity of the obtained correction value in patients having sulcus fixation.Department of Ophthalmology, Saiseikai Kurihashi Hospital, Saitama, Japan.The postoperative refraction, virtual in-the-bag IOL power for the same postoperative refraction, predicted refraction with an increase in in-the-bag IOL power by 0.50 to 2.00 diopters (D), and effective lens position (ELP) in 30 eyes with a normal axial length after sulcus fixation were retrospectively evaluated. The results were confirmed prospectively in 16 other eyes requiring sulcus fixation.A significant myopic shift (mean 0.78 D +/- 0.47 [SD]) from the predicted refraction occurred after sulcus fixation (P<.0001). The mean virtual IOL power was 1.11 +/- 0.67 D higher than that of the actual sulcus-fixated IOL. The difference between the predicted refraction and the refraction after sulcus fixation was less when the difference in IOL power was 1.00 D. The ELP of the IOLs was 0.75 mm shorter using A-mode measurement and 0.62 mm shorter using back calculation from the theoretical formula with sulcus fixation than with in-the-bag fixation. In the prospective study, there was a small refractive error (mean -0.06 +/- 0.19 D) when the planned IOL power was reduced by 1.00 D for sulcus fixation.For sulcus fixation in eyes with a normal axial length, the IOL power should be 1.00 D less than the power for in-the-bag fixation.
机译:为了确定对预测的人工晶状体(IOL)进行沟内固定的适当矫正,并前瞻性地评估所获得的矫正值在有沟内固定的患者中的有效性。日本ita玉县Saiseikai Kurihashi医院眼科。相同眼屈光度的袋内人工晶状体屈光度,预计屈光度随袋内人工晶状体屈光度增加0.50至2.00屈光度(D)以及30眼正常眼轴的有效晶状体位置(ELP)对沟内固定后的患者进行回顾性评估。结果在另外16颗需要进行眼沟固定的眼睛中得到了前瞻性的确认。预测的屈光度在眼沟固定后发生了显着的近视移位(平均0.78 D +/- 0.47 [SD])(P <.0001)。虚拟虚拟IOL的平均功率比实际的沟固定IOL高1.11 +/- 0.67D。当IOL屈光力之差为1.00 D时,预测屈光度与沟固定后屈光度之间的差异较小。使用A模式测量,IOL的ELP缩短0.75 mm,而根据理论公式,使用反向计算,ILP的ELP缩短0.62 mm槽内固定比袋内固定。在前瞻性研究中,当将计划的IOL屈光力降低1.00 D进行眼沟内固定时,屈光误差较小(平均-0.06 +/- 0.19 D)。对于正常轴长的眼内眼沟固定,IOL屈光度应比袋内固定的功率小1.00D。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号