Sagittal curvature of total knee replacements predicts in vivo kinematics.

机译:全膝关节置换的矢状曲率可预测体内运动学。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: It is known that in vivo kinematics after total knee replacement is influenced by the design of the implant. The goal of this study was to show that the sagittal curvature of two different knee prostheses differing in geometric design predicts their in vivo motion behavior. METHODS: Three-dimensional tibio-femoral displacements of two prosthesis designs (single radius vs. dual radius) were measured during knee extension under weight bearing conditions by in vivo video fluoroscopy. Finite helical axes were computed to represent the tibio-femoral motions. Angular deviation alpha and the spatial localization deviation delta were used to characterize the motions. Angular deviation is the angle between each incremental finite helical axis and the medio-lateral axis of the femoral component of the prosthesis. The spatial localization deviation is the distance between each finite helical axis and the center of the femoral component of the prosthesis. Statistical comparisons were performed using the median and the interquartile range of the angular deviation and the spatial localization deviation. FINDINGS: The single-radius design showed finite helical axes concentrated at a single axis near to the medio-lateral axis of the femoral component. The angular and spatial localization deviation of the dual radius design were larger compared to the single radius design, exhibiting finite helical axes varying between two axes. INTERPRETATION: Video fluoroscopy in combination with finite helical axis analysis proved to be suitable methods to evaluate the in vivo kinematical behavior of total knee arthroplasty, which can be useful for implant designers. Knowledge of in vivo kinematics can also provide surgeons with more background information about the total knee arthroplasty models they implant.
机译:背景:众所周知,全膝关节置换后的体内运动学受植入物设计的影响。这项研究的目的是表明,在几何设计上不同的两个不同的膝假体的矢状弯曲可以预测其体内运动行为。方法:在体内负重的条件下,通过体内视频荧光检查法测量了两种假体设计的三维股骨位移(单radius对双radius)。计算有限的螺旋轴以代表胫股运动。角偏差α和空间定位偏差δ用于表征运动。角偏差是每个增量有限螺旋轴与假体股骨组件的中外侧轴之间的角度。空间定位偏差是每个有限螺旋轴与假体股骨组件中心之间的距离。使用角度偏差和空间定位偏差的中位数和四分位间距进行统计比较。结果:单半径设计显示有限的螺旋轴集中在靠近股骨组件的中外侧轴的单个轴上。与单半径设计相比,双半径设计的角度和空间定位偏差更大,表现出有限的螺旋轴在两个轴之间变化。解释:荧光透视结合有限螺旋轴分析被证明是评估全膝关节置换术体内运动学行为的合适方法,对植入物设计者可能是有用的。体内运动学知识还可以为外科医生提供有关他们植入的全膝关节置换模型的更多背景信息。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号