首页> 外文期刊>Journal of Biomechanics >Direct comparison of calculated hip joint contact forces with those measured using instrumented implants. An evaluation of a three-dimensional mathematical model of the lower limb.
【24h】

Direct comparison of calculated hip joint contact forces with those measured using instrumented implants. An evaluation of a three-dimensional mathematical model of the lower limb.

机译:将计算出的髋关节接触力与使用仪器植入物测得的直接比较。下肢的三维数学模型的评估。

获取原文
获取原文并翻译 | 示例
           

摘要

Characterisation of hip joint contact forces is essential for the definition of hip joint prosthesis design requirements. In vivo hip joint contact force measurements have been made using instrumented hip joint prostheses. However, to allow determination of the range of values of joint contact force and their directions relative to anatomical structures in a range of subject groups sufficient to form an agreed data base it is necessary to adopt a different approach without the use of an implanted transducer. The use of mathematical models of the lower limb to examine the forces in soft tissues and at the joints has provided valuable insight into internal loading conditions. Several authors have proposed mathematical musculo-skeletal models. However, there have been only limited attempts at validation of these models. It is possible to use the results of in vivo force measurements from instrumented prostheses to validate the results calculated using the mathematical models. In this study two subjects with instrumented hip joint prostheses were studied. Forces at the hip joints were calculated using a three-dimensional model of the leg. Walking at slow, normal and fast speeds (0.97-2.01m/s), weight transfer from two to one leg and back again, and sit to stand were studied. Direct comparisons were made between the 'gold standard' measured hip joint contact forces and the calculated forces. There was general agreement between the calculated and measured forces in both pattern and magnitude. There were, however, discrepancies. Reasons for these differences in results are discussed and possible model developments suggested.
机译:髋关节接触力的表征对于定义髋关节假体设计要求至关重要。已经使用仪器化的髋关节假体进行了体内髋关节接触力的测量。但是,为了确定关节接触力值的范围及其相对于足以形成一致数据库的主题组范围内的解剖结构的方向,必须采用不同的方法而不使用植入的换能器。使用下肢的数学模型检查软组织中和关节处的力,可以深入了解内部负荷情况。一些作者提出了数学上的肌肉骨骼模型。但是,只有有限的尝试来验证这些模型。可以使用器械假体的体内力测量结果来验证使用数学模型计算出的结果。在这项研究中,研究了两个带有髋关节假体的受试者。使用腿部的三维模型计算髋关节的力。研究了以慢速,正常和快速速度(0.97-2.01m / s)行走,从两条腿转移到另一只腿并再次向后转移并坐着站立的情况。在“金标准”测得的髋关节接触力和计算出的力之间进行了直接比较。在模式和大小上,计算出的力和测量出的力之间普遍达成一致。但是,存在差异。讨论了这些结果差异的原因,并提出了可能的模型开发建议。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号