首页> 外文期刊>Journal of Biomechanics >In vitro comparison of bone deformation measured with surface and staple mounted strain gauges
【24h】

In vitro comparison of bone deformation measured with surface and staple mounted strain gauges

机译:用表面应变计和钉书钉应变计测量的骨变形的体外比较

获取原文
获取原文并翻译 | 示例
           

摘要

Chicken tibiae were chosen as a model for human second metatarsals. Local surface bone deformation in a 4-point bending configuration was measured in vitro by both strain gauge instrumented staples and strain gauges bonded to the bone's cortical surface. A series of staple bridge dimensions (0.5, 0.6, 0.8 and 1.0 mm) was compared to test for staple influence on bone characteristics and greatest measurement validity and reliability. Thicker staple inhibition of bone deformation was the greatest but differences to thinner staples were not statistically significant (p>0.05). All staples except 0.5 mm had maximum deviations from linearity less than 1%. The 1.0 mm staple had an R super(2) value of 0.992 plus or minus 0.006 plotted against the 4-point bending input force and 0.994 plus or minus 0.002 plotted against the surface strain gauge signal. The mean intraclass correlation coefficients (ICC) calculated with four input forces (30, 60, 90 and 120 N) and for loading and unloading conditions for the 0.5, 0.6, 0.8 and 1.0 mm staples were 0.75, 0.83, 0.87 and 0.92, respectively. Finally, the differences in slope of the staple strain gauge signal plotted against surface strain gauge signal between input force loading and unloading conditions (0.32), and between input compression and tension conditions (0.79) was least for the 1.0 mm staple which also resulted in the lowest standard deviations. These results suggested the appropriateness of the 1.0 mm staple for in vivo application.
机译:选择胫骨小鸡作为人类第二meta骨的模型。在体外通过测量应变仪的钉书钉和粘合到骨皮质表面的应变仪测量了4点弯曲配置下的局部表面骨变形。比较了一系列吻合钉桥的尺寸(0.5、0.6、0.8和1.0 mm),以测试吻合钉对骨骼特性的影响以及最大的测量有效性和可靠性。较厚的吻合钉对骨变形的抑制作用最大,但与较薄的吻合钉的差异无统计学意义(p> 0.05)。除0.5毫米外,所有订书钉的线性最大偏差均小于1%。 1.0毫米订书钉的R super(2)值相对于4点弯曲输入力绘制为0.992正负0.006,相对于表面应变仪信号绘制为0.994正负0.002。在四个输入力(30、60、90和120 N)下以及在0.5、0.6、0.8和1.0 mm订书钉的装卸条件下计算出的组内平均相关系数(ICC)分别为0.75、0.83、0.87和0.92 。最后,对于1.0 mm的订书钉,钉书钉应变仪信号相对于表面应变仪信号的斜率与输入力加载和卸载条件(0.32)之间以及输入压缩和拉伸条件(0.79)之间的差异最小。最低标准偏差。这些结果表明1.0mm缝合钉适合于体内应用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号