Telemeterized internal spinal fixation devices were implanted in a patient with degenerative instability and a narrow spinal canal in order to measure the fixator loads during daily activities. Anterior interbody fusion was performed three weeks later. During walking, the typical maximum flexion bending moments were 10 N m in the left and 5 N m in the right fixator. On removal of the implants three months later, a fatigue fracture was found not on the high loaded left side but in the upper right pedicle screw. The crack started on the caudal side of the cross-sectional area and progressed cranially. Upper vertebral tilting in the sagittal plane must have caused the screw breakage. This would probably have been prevented by a more posteriorly placed bone graft.
机译:遥测式内部脊柱内固定装置被植入具有退行性不稳定和狭窄的椎管的患者中,以便在日常活动中测量固定器的负荷。三周后进行前路椎间融合术。在行走过程中,典型的最大屈曲弯矩在左侧为10 N m,在右侧为5 N m。三个月后取出植入物时,发现疲劳断裂不是在高负荷左侧,而是在右上椎弓根螺钉上。裂纹从横截面的尾侧开始并逐渐扩展。矢状平面中的上椎骨倾斜必定导致了螺钉断裂。可以通过在后面放置更多的骨移植物来防止这种情况。