首页> 外文期刊>Journal of biomaterials applications >Stent fabric fatigue of grafts supported by Z-stents versus ringed stents: An in vitro buckling test
【24h】

Stent fabric fatigue of grafts supported by Z-stents versus ringed stents: An in vitro buckling test

机译:Z型支架与环形支架支撑的移植物的支架织物疲劳:体外屈曲测试

获取原文
获取原文并翻译 | 示例
           

摘要

Stent-grafts externally fitted with a Z-shaped stents were compared to devices fitted with ringed stents in an in vitro oscillating fatigue machine at 200 cycles per minute and a pressure of 360 mmHg for scheduled durations of up to 1 week. The devices fitted with Z-stents showed a considerably lower endurance limit to buckling compared to the controls. The contact between the apexes of adjacent Z-stents resulted in significant damage to the textile scaffolds and polyester fibers due to the sharp angle of the Z-stents. The ringed stents did not cause any fraying in the textile scaffolds.
机译:在体外振荡疲劳仪中,将外部装有Z形支架的支架移植物与装有环形支架的装置进行了比较,该设备在体外振荡疲劳机中以每分钟200个循环的速度和360 mmHg的压力进行了长达1周的计划时间。与对照组相比,装有Z形支架的设备对屈曲的承受力极限要低得多。相邻Z形支架顶点之间的接触由于Z形支架的锐角而导致对纺织品支架和聚酯纤维的重大损坏。环状支架不会在纺织品支架上造成任何磨损。