首页> 外文期刊>The journal of asthma >Comparison of the ATS versus EU Mini Wright peak flow meter in normal volunteers.
【24h】

Comparison of the ATS versus EU Mini Wright peak flow meter in normal volunteers.

机译:正常志愿者中ATS与EU Mini Wright峰值流量计的比较。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVES: The American Thoracic Society (ATS) and European Union (EU) have precise and accurate Mini Wright peak flow meters. The purpose of this investigation was to compare both 1) for accuracy using a pneumotachometer, 2) in volunteers to determine whether they are interchangeable, and 3) to spirometrically predicted peak flows. METHODS: Lab testing: A pneumotachometer was connected in series with each peak flow meter and varying flows pushed through both meters for comparison. Human subjects: Nonsmoking adult volunteers did three standing peak flows. The order of peak flow meter used was random. The best of three efforts was used for analysis. The t-test, concordance correlation coefficient (CCC), Deming regression, and Bland-Altman plot were the analytic strategies used to determine agreement. Peak flow results were compared to spirometrically predicted values. RESULTS: Fifty-seven volunteers, average age 37 +/- 12 years and mean BMI 24.9 +/- 2.5 years, were included. The average peak flows were different at 541 +/- 114 and 526 +/- 112 L/min for the ATS and EU meters, respectively (p < .01). Both peak flow meter values were significantly different than spirometrically predicted values of 483 +/- 86 L/min (p < .01). The CCC was 0.98 (0.97-0.99) and regression revealed a slope and y-intercept consistent with 1 and 0, respectively. The Bland-Altman plot revealed no increase in scatter of values over the range of peak flows versus the difference with a mean bias of 15 +/- 15 L/min. Laboratory testing revealed that the ATS and EU peak flow meters read 3.0 +/- 2.1% above and -2.0 +/- 1.5% below the comparison pneumotachometer, respectively. The pneumotachometer comparison was significantly different for both meters at p < .01, paired t-test. CONCLUSIONS: The ATS peak flow meter reads 2.8% higher than the EU peak flow meter across a range of flows. Both meters have similar accuracy with a different bias compared with a pneumotachometer. Finally, both peak flow meters read slightly and significantly higher than spirometrically derived peak flows. Therefore, the peak flow meters are not interchangeable and both may obtain slightly higher values than those determined using current spirometrically derived prediction equations.
机译:目标:美国胸科学会(ATS)和欧盟(EU)拥有精确精确的Mini Wright峰值流量计。这项研究的目的是比较1)使用气压计的准确性,2)志愿者确定它们是否可互换以及3)与肺活量预测的峰值流量进行比较。方法:实验室测试:将气动转速计与每个峰值流量计串联连接,并通过两个仪表推动不同流量进行比较。人类受试者:不吸烟的成年志愿者表现出三个站立高峰。峰值流量计的使用顺序是随机的。三项努力中最好的一项用于分析。 t检验,一致性相关系数(CCC),Deming回归和Bland-Altman图是用于确定一致性的分析策略。将峰值流量结果与肺功能测定的预测值进行比较。结果:五十七名志愿者,平均年龄37 +/- 12岁,平均BMI 24.9 +/- 2.5岁。对于ATS和EU流量计,平均峰值流量分别为541 +/- 114和526 +/- 112 L / min(p <0.01)。流量计的两个峰值均与483 +/- 86 L / min的肺功能测定法预测值显着不同(p <0.01)。 CCC为0.98(0.97-0.99),回归显示斜率和y截距分别与1和0一致。 Bland-Altman图显示,在峰值流量范围内,值的散布没有增加,而平均偏差为15 +/- 15 L / min。实验室测试表明,ATS和EU峰值流量计的读数分别比比较式气压计高3.0 +/- 2.1%和-2.0 +/- 1.5%。配对的t检验在p <.01时,两个仪表的气压计比较存在显着差异。结论:在一系列流量范围内,ATS峰值流量计的读数比EU峰值流量计高2.8%。与气动转速表相比,两种仪表的精度相似,偏置不同。最终,两个峰值流量计的读数都略微大于显着高于通过呼吸计量得出的峰值流量。因此,峰值流量计是不可互换的,并且两者都可以获得比使用当前的呼吸测定法得出的预测方程式确定的峰值稍高的值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号