In this study, an electrochemical method was developed using flow injection analysis and amperometric detection with screen-printed electrodes for iodide determination as a tracer in seawater reservoirs. The amperometric method resulted in a good linear correlation coefficient (r = 0.9966) with a linear response over a concentration range of 1-10 mg L ~(-1), a limit of detection of 0.2 mg L ~(-1) and a quantification of 0.6 mg L ~(-1). In addition, the method has been shown to be precise, linear and homoscedastic. The recovery for the amperometric method was 100 ± 3 %, and the recovery for the produced water samples fortified with iodide was 102 ± 2 %. The amperometric method did not show interference from other anions. Thus, the proposed method is promising for analysing the iodide content of seawater reservoir samples with simple and inexpensive methodologies compared with ion chromatography techniques.
机译:在这项研究中,开发了一种电化学方法,该方法使用了流动注射分析和安培检测,并使用丝网印刷电极作为海水中示踪剂的碘化物测定。安培法产生了良好的线性相关系数(r = 0.9966),在1-10 mg L〜(-1)的浓度范围内具有线性响应,检出限为0.2 mg L〜(-1),并且定量为0.6 mg L〜(-1)。此外,该方法已被证明是精确,线性和均等的。安培法的回收率为100±3%,碘化物强化的采出水样品的回收率为102±2%。安培法未显示出其他阴离子的干扰。因此,与离子色谱技术相比,所提出的方法有望以简单且廉价的方法来分析海水储层样品中的碘含量。