首页> 外文期刊>Journal of Applied Crystallography >A microfocus X-ray tube used with focusing collimators
【24h】

A microfocus X-ray tube used with focusing collimators

机译:用于聚焦准直器的微焦点X射线管

获取原文
获取原文并翻译 | 示例
           

摘要

The design and performance are described of a low-power copper-target X-ray tube with an elongated focus foreshortened to a diameter between 10 and 20 mu m. The tube is designed for use with a specularly reflecting ellipsoidal mirror placed between 10 and 20 mm from the source and focusing the X-rays onto the sample, which may be 600 mm distant. The design maximizes the solid angle of collection of the emitted X-rays and thus achieves a high intensity at the sample. The electron beam, of circular cross section, from the gun is focused by an axial magnetic lens and then drawn out by a quadrupole lens to form an elongated spot on the target which is viewed at a small take-off angle. The Limits imposed on the maximum intensity at the specimen by aberrations in the lens and by the permissible power dissipation in the target are discussed. The observed intensity is in reasonable agreement with the calculated intensity; it is comparable with that achieved with nonfocusing X-ray optics used with conventional X-ray tubes operated at a power more than 100 times as great. [References: 18]
机译:描述了低功率铜靶X射线管的设计和性能,该靶的细长焦点缩短到10至20μm之间。该管设计为与镜面反射椭球面镜配合使用,该镜面距光源10到20毫米,并将X射线聚焦到距离600毫米的样品上。该设计最大程度地提高了发射X射线的收集立体角,从而在样品处实现了高强度。来自电子枪的圆形横截面的电子束由轴向磁透镜聚焦,然后由四极透镜引出,以在靶子上形成细长点,以较小的起飞角观察即可。讨论了由镜片上的像差和目标中允许的功率耗散对试样最大强度施加的极限。观察到的强度与计算出的强度在合理范围内;它可以与以100倍以上的功率工作的常规X射线管使用的非聚焦X射线光学器件相媲美。 [参考:18]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号