首页> 外文期刊>Journal of Applied Crystallography >SMEARING EFFECTS INTRODUCED BY THE RESPONSE FUNCTION OF POSITION-SENSITIVE GAS DETECTORS IN SAXS EXPERIMENTS
【24h】

SMEARING EFFECTS INTRODUCED BY THE RESPONSE FUNCTION OF POSITION-SENSITIVE GAS DETECTORS IN SAXS EXPERIMENTS

机译:SAXS实验中位置敏感型气体检测器的响应功能引入的测量效果

获取原文
获取原文并翻译 | 示例

摘要

The response of gas-filled position-sensitive detectors consists not only of the main Gaussian response (FWHM similar to 0.1 mm) but also of a linear exponential term measurable up to 7 mm from the excitation, superimposed on an incoherent background. This effect depends on gas type and pressure and is due to fluorescence of the detection gas. The consequences of this effect on some standard experimental methods used in small-angle X-ray scattering (SAXS), i.e. Porod limit and Guinier radii determination, are investigated. [References: 14]
机译:充气的位置敏感探测器的响应不仅由主要的高斯响应(FWHM近似于0.1 mm)组成,而且还包括一个线性指数项,可测量到距激发不超过7 mm的位置,并叠加在非相干背景上。该影响取决于气体类型和压力,并且是由于检测气体的荧光所致。研究了这种影响对小角度X射线散射(SAXS)中使用的某些标准实验方法(即Porod极限和Guinier半径测定)的影响。 [参考:14]