Intracoronary optical coherence tomography and histology of overlapping everolimus-eluting bioresorbable vascular scaffolds in a porcine coronary artery model: The potential implications for clinical practice

机译:猪冠状动脉模型中重叠依维莫司洗脱生物可吸收性血管支架的冠状动脉内光学相干断层扫描和组织学:对临床实践的潜在影响

获取原文
获取原文并翻译 | 示例
           

摘要

Objectives: This study sought to assess the vascular response of overlapping Absorb stents compared with overlapping newer-generation everolimus-eluting metallic platform stents (Xience V [XV]) in a porcine coronary artery model. Background: The everolimus-eluting bioresorbable vascular scaffold (Absorb) is a novel approach to treating coronary lesions. A persistent inflammatory response, fibrin deposition, and delayed endothelialization have been reported with overlapping first-generation drug-eluting stents. Methods: Forty-one overlapping Absorb and overlapping Xience V (XV) devices (3.0 × 12 mm) were implanted in the main coronary arteries of 17 nonatherosclerotic pigs with 10% overstretch. Implanted coronary arteries were evaluated by optical coherence tomography (OCT) at 28 days (Absorb n = 11, XV n = 7) and 90 days (Absorb n = 11, XV n = 8), with immediate histological evaluation following euthanasia at the same time points. One animal from each time point was evaluated with scanning electron microscopy alone. A total of 1,407 cross sections were analyzed by OCT and 148 cross sections analyzed histologically. Results: At 28 days in the overlap, OCT analyses indicated 80.1% of Absorb struts and 99.4% of XV struts to be covered (p < 0.0001), corresponding to histological observations of struts with cellular coverage of 75.4% and 99.6%, respectively (p < 0.001). Uncovered struts were almost exclusively related to the presence of "stacked" Absorb struts, that is, with a direct overlay configuration. At 90 days, overlapping Absorb and overlapping XV struts demonstrated >99% strut coverage by OCT and histology, with no evidence of a significant inflammatory process, and comparable % volume obstructions. Conclusions: In porcine coronary arteries implanted with overlapping Absorb or overlapping XV struts, strut coverage is delayed at 28 days in overlapping Absorb, dependent on the overlay configuration of the thicker Absorb struts. At 90 days, both overlapping Absorb and overlapping XV have comparable strut coverage. The implications of increased strut thickness may have important clinical and design considerations for bioresorbable platforms.
机译:目的:本研究旨在评估在猪冠状动脉模型中,与重叠的新一代依维莫司洗脱金属平台支架(Xience V [XV])相比,重叠的Absorb支架的血管反应。背景:依维莫司洗脱生物可吸收血管支架(Absorb)是一种治疗冠状动脉病变的新方法。据报道,重叠的第一代药物洗脱支架具有持续的炎症反应,纤维蛋白沉积和内皮化延迟。方法:将四十一只重叠的Absorb和Xience V(XV)重叠装置(3.0×12毫米)植入17只非动脉粥样硬化猪的主冠状动脉中,其伸展度为10%。通过光学相干断层扫描(OCT)在28天(吸收n = 11,XV n = 7)和90天(吸收n = 11,XV n = 8)评估植入的冠状动脉,并在相同的安乐死后立即进行组织学评估时间点。在每个时间点对一只动物进行单独的扫描电子显微镜评估。通过OCT分析了总共1,407个横截面,并通过组织学分析了148个横截面。结果:在重叠的28天中,OCT分析表明,有80.1%的Absorb支撑物和99.4%的XV支撑物被覆盖(p <0.0001),分别对应于其细胞覆盖率为75.4%和99.6%的支撑物的组织学观察( p <0.001)。裸露的撑杆几乎完全与“堆叠式” Absorb撑杆的存在有关,即具有直接覆盖结构。在90天时,OCT和组织学证实重叠的Absorb和XV支撑物覆盖率> 99%,没有明显的炎症过程,也没有体积阻塞的百分比。结论:在猪冠状动脉中植入重叠的Absorb支杆或XV重叠的支杆,重叠Absorb支杆的覆盖时间会延迟28天,这取决于较厚的Absorb支杆的覆盖结构。在90天时,重叠的Absorb和重叠的XV的支撑杆覆盖率均相当。支撑杆厚度增加的含义对于生物可吸收平台可能具有重要的临床和设计考虑。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号