Assessment of different techniques for subcutaneous glucose monitoring in Type 1 diabetic patients during 'real-life' glucose excursions

机译:在“现实生活”中的葡萄糖偏移期间评估1型糖尿病患者皮下葡萄糖监测的不同技术

获取原文
获取原文并翻译 | 示例
           

摘要

Aims To compare the accuracy of two marketed subcutaneous glucose monitoring devices (Guardian RT, GRT; GlucoDay S, GDS) and standard microdialysis (CMA60; MD) in Type 1 diabetic patients.Methods Seven male Type diabetic patients were investigated over a period of 26 h simulating real-life meal glucose excursions. Catheters of the three systems were inserted into subcutaneous adipose tissue of the abdominal region. For MD, interstitial fluid was sampled at 30- to 60-min intervals for offline glucose determination. Reference samples were taken at 15- to 60-min intervals. All three systems were prospectively calibrated to reference. Median differences, median absolute relative differences (MARD), median absolute differences (MAD), Bland-Altman plot and Clark Error Grid were used to determine accuracy.
机译:目的比较两种市售的皮下葡萄糖监测设备(Guardian RT,GRT; GlucoDay S,GDS)和标准微透析(CMA60; MD)在1型糖尿病患者中的准确性。方法在26名患者中调查了7名男性糖尿病患者h模拟现实生活中的膳食葡萄糖偏移。将三个系统的导管插入腹部区域的皮下脂肪组织中。对于MD,以30到60分钟的间隔采样组织液以进行离线葡萄糖测定。每隔15至60分钟采集一次参考样品。所有三个系统均经过前瞻性校准以供参考。中位数差异,中位数绝对相对差异(MARD),中位数绝对差异(MAD),Bland-Altman图和Clark错误网格用于确定准确性。