Effective dose span of ten different cone beam CT devices

机译:十种不同的锥形束CT设备的有效剂量跨度

获取原文
获取原文并翻译 | 示例
           

摘要

Objectives: Evaluation and reduction of dose are important issues. Since cone beam CT (CBCT) has been established now not just in dentistry, the number of acquired examinations continues to rise. Unfortunately, it is very difficult to compare the doses of available devices on the market owing to different exposition parameters, volumes and geometries. The aim of this study was to evaluate the spans of effective doses (EDs) of ten different CBCT devices. Methods: 48 thermoluminescent dosemeters were placed in 24 sites in a RANDO? head phantom. Protocols with lowest exposition parameters and protocols with highest exposition parameters were performed for each of the ten devices. The ED was calculated from the measured energy doses according to the International Commission on Radiological Protection 2007 recommendations for each protocol and device, and the statistical values were evaluated afterwards. Results: The calculation of the ED resulted in values between 17.2 μSv and 396 μSv for the ten devices. The mean values for protocols with lowest and highest exposition parameters were 31.6 μSv and 209 μSv, respectively. Conclusions: It was not the aim of this study to evaluate the image quality depending on different exposition parameters but to define the spans of EDs in which different CBCT devices work. There is a wide span of ED for different CBCT devices depending on the selected exposition parameters, required spatial resolution and many other factors.
机译:目标:评估和减少剂量是重要的问题。由于现在不仅在牙科领域已经建立了锥形束CT(CBCT),因此获得的检查数量还在继续增加。不幸的是,由于不同的曝光参数,体积和几何形状,很难比较市场上可用设备的剂量。这项研究的目的是评估十种不同CBCT设备的有效剂量(ED)跨度。方法:将48台热发光剂量计放置在RANDO?的24个位置中。头幻影。对十个设备中的每一个都执行了具有最低暴露参数的协议和具有最高暴露参数的协议。根据国际放射防护委员会2007年建议,针对每种方案和设备,由测得的能量剂量计算出ED,然后评估统计值。结果:十个器件的ED计算得出值在17.2μSv和396μSv之间。具有最低和最高暴露参数的方案的平均值分别为31.6μSv和209μSv。结论:本研究的目的不是根据不同的曝光参数评估图像质量,而是定义不同CBCT设备在其中工作的ED的跨度。根据所选的曝光参数,所需的空间分辨率和许多其他因素,不同的CBCT设备的ED范围很广。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号