Objectives: This study compares the effective dose for different fields of view (FOVs), resolutions and X-ray parameters from two cone beam CT units: the KaVo 3D (three-dimensional) eXam and the KaVo Pan eXam Plus 3D (KaVo Dental, Biberach, Germany). Methods: Measurementswere made using thermoluminescent dosemeter chips in a radiation analog dosimetry head and neck phantom. The calculations of effective doses are based on the ICRP 60 and ICRP 103 recommendations of the International Commission on Radiological Protection. Results: Effective doses from the 3D eXam ranged between 32.8 μSv and 169.8 μSv, and for the Pan eXam Plus effective doses ranged between 40.2 μSv and 183.7 μSv; these were measured using ICRP 103 weighting factors in each case. The increase in effective dose between ICRP 60 and ICRP 103 recommendations averaged 157% for all measurements. Conclusions: Effective doses can be reduced significantly with the choice of lower resolutions and mAs settings as well as smaller FOVs to avoid tissues sensitive to radiation being inside the direct beam. Larger FOVs do not necessarily lead to higher effective doses.
机译:目的:本研究比较了两个锥束CT单元:KaVo 3D(三维)eXam和KaVo Pan eXam Plus 3D(KaVo Dental,两种)的不同视场(FOV),分辨率和X射线参数的有效剂量。德国比伯拉赫)。方法:使用热辐射剂量计芯片在辐射模拟剂量学头颈部体模中进行测量。有效剂量的计算基于国际放射防护委员会的ICRP 60和ICRP 103建议。结果:3D eXam的有效剂量范围为32.8μSv至169.8μSv,Pan eXam Plus的有效剂量范围为40.2μSv至183.7μSv。在每种情况下,均使用ICRP 103加权因子来测量这些值。对于所有测量,ICRP 60和ICRP 103建议之间的有效剂量增加平均为157%。结论:通过选择较低的分辨率和mAs设置以及较小的FOV,可以显着降低有效剂量,以避免对辐射敏感的组织位于直射光束内。较大的FOV不一定会导致较高的有效剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号