首页> 外文期刊>Vascular >Aorfix stent graft for abdominal aortic aneurysms reduces the risk of proximal type 1 endoleak in angulated necks: bench-test study.
【24h】

Aorfix stent graft for abdominal aortic aneurysms reduces the risk of proximal type 1 endoleak in angulated necks: bench-test study.

机译:腹主动脉瘤的Aorfix支架移植物可减少颈椎近端1型内漏的风险:基准试验研究。

获取原文
获取原文并翻译 | 示例
           

摘要

Neck angulation (NA) is an important risk factor for type 1 proximal endoleaks following stenting of abdominal aortic aneurysms. The Aorfix (Lombard Medical, Oxon, UK) is a new flexible stent graft designed to overcome this issue. The aim of this study was to compare the endoleak flow rate (EFR) in relation to NA between the Aorfix and other manufactured stent grafts. A flow model with silicone proximal and distal necks was used. EFRs corresponding to 10 neck angles between 0 and 70 degrees were measured. Eight stent grafts were tested: Aorfix, Ancure (Guidant, Indianapolis, IN), Powerlink (Endologix, Irvine, CA), AneuRx (Medtronic, Sunnyvale, CA), Excluder (W.L. Gore & Associates, Flagstaff, AZ), Zenith and Zenith-Flex (Cook Inc., Bloomington, IN), and Lifepath (Edwards Lifesciences, Irvine, CA). For all stent grafts except the Aorfix, the EFR was greater than at baseline for NA >or= 30 degrees (p < .01). The EFR at NA >or= 30 degrees was lower with the Aorfix compared with the other stent grafts (p <.01). NA had no influence on the EFR with the Aorfix. The Aorfix may decrease the incidence of proximal type 1 endoleak in patients with a severely angulated aortic neck.
机译:颈动脉成角(NA)是在腹主动脉瘤支架置入术后1型近端内漏的重要危险因素。 Aorfix(英国Oxon的Lombard Medical)是一种新型的柔性支架移植物,旨在克服这一问题。这项研究的目的是比较Aorfix和其他制造的覆膜支架之间相对于NA的内漏流速(EFR)。使用带有硅胶近端和远端颈部的流动模型。测量了与0到70度之间的10个颈角相对应的EFR。测试了八种覆膜支架:Aorfix,Ancure(印第安那州印第安纳波利斯的Guidant),Powerlink(加利福尼亚州欧文市的Endologix),AneuRx(加利福尼亚州桑尼维尔的美敦力公司),Excluder(WL Gore&Associates,弗拉格斯塔夫,亚利桑那州),Zenith和Zenith -Flex(Cook Inc.,布卢明顿,印第安那州)和Lifepath(Edwards Lifesciences,尔湾,加利福尼亚州)。对于NA>或= 30度的除Aorfix以外的所有支架移植物,EFR均大于基线(p <.01)。与其他覆膜支架相比,Aorfix的NA>或= 30度时的EFR较低(p <.01)。 NA对使用Aorfix的EFR没有影响。 Aorfix可降低主动脉颈严重弯曲患者近端1型内漏的发生率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号