Radiography and biomechanics of sixth and seventh cervical vertebrae segments after disc fenestration and after insertion of an intervertebral body spacer: a canine cadaveric study.

机译:椎间盘开窗后和插入椎体间隔器后第六和第七颈椎节段的放射照相和生物力学:犬尸体研究。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Objectives: To study the radiographic characteristics and the biomechanical properties of the sixth and seventh cervical (C6-C7) vertebral motion unit (VMU) with an intact disc, after disc fenestration, and after placement of an intervertebral body spacer (IVBS). Methods: Six cadaveric C6-C7 VMU were retrieved from six Greyhound cadavers. Each VMU was loaded at 3 Nm of torque sequentially in flexion, extension, and in right and left lateral bending. The range-of-motion (ROM) was measured with a Zebris 3DReg. system. The intervertebral disc cross-sectional area was measured on lateral and ventro-dorsal radiographs. Biomechanical testing and radiographic measurements were performed with an intact disc, after disc fenestration, and after IVBS placement. Data were reported as mean+or-SD. Results: The intervertebral disc cross-sectional area was significantly decreased after disc fenestration and increased after IVBS placement, but remained significantly smaller than the area of intact disc in some of the tested conditions. The ROM with an intact disc, after disc fenestration and after IVBS placement, in flexion were 11.5 degrees +or-1.0, 15.2 degrees +or-2.3, and 10.9 degrees +or-4.7, respectively, and in extension were 15.6 degrees +or-3.7, 24.7 degrees +or-6.2, 21.9 degrees +or-4.0, respectively. There was a significant increase in extension ROM after disc fenestration. Intervertebral body spacer placement significantly decreased ROM in flexion but ROM in extension was not different from disc fenestration. No significant changes in lateral bending ROM were detected. Clinical significance: The use of an IVBS reduced disc space collapse but did not restore stability of the VMU to normal values in extension after cervical disc fenestration.
机译:目的:研究椎间盘开窗后,放置椎间间隔物(IVBS)后,具有完整椎间盘的第六和第七颈椎(C6-C7)椎体运动单元(VMU)的放射学特征和生物力学特性。方法:从六个灵狮尸体中检索出六个尸体C6-C7 VMU。每个VMU依次以3 Nm的扭矩加载,进行弯曲,伸展以及左右弯曲。使用Zebris 3DReg测量运动范围(ROM)。系统。椎间盘的横截面积是在侧面和腹背片上测量的。在椎间盘开窗后和放置IVBS后,对完整的椎间盘进行生物力学测试和射线照相测量。数据报道为平均值±SD。结果:椎间盘开窗后,椎间盘的横截面积显着减少,而放置IVBS后,椎间盘的横截面积增加,但在某些测试条件下仍显着小于完整椎间盘的面积。具有完整椎间盘的ROM,在椎间盘开窗后和IVBS放置后,屈曲度分别为11.5度或-1.0、15.2度+或-2.3和10.9度+或-4.7,延伸时为15.6度+或-。 -3.7、24.7度或-6.2、21.9度或-4.0。光盘开窗后,扩展ROM显着增加。椎体间隔垫片的放置显着降低了ROM的屈曲,但延伸的ROM与椎间盘开窗术没有区别。没有检测到横向弯曲ROM的显着变化。临床意义:使用IVBS可以减少椎间盘间隙塌陷,但不能将VMU的稳定性恢复到宫颈椎间盘开窗后延伸的正常值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号