首页> 外文期刊>Tribology International >Effect of surface treatments on fretting fatigue damage of biomedical titanium alloys
【24h】

Effect of surface treatments on fretting fatigue damage of biomedical titanium alloys

机译:表面处理对生物医学钛合金微动疲劳损伤的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Fretting fatigue is an adhesive wear damage caused by tangential micromotion under normal force at contact areas. It is observed along the contact points of hip implants and bone plates. Surface-modified biomedical titanium alloys offer better resistance against fretting damage. PVD TiN coatings and plasma nitriding have proved effective in minimizing friction and delaying the failure of materials. In the present study, attempt has been made to explain the fretting fatigue failure mechanism sequence of PVD TiN-coated and plasma-nitrided Ti-6Al-4V and Ti-6Al-4V couple through friction measurement and microscopic examination.
机译:微动疲劳是指在接触区域的法向力下,由于切向微运动引起的粘合剂磨损损坏。沿着髋关节植入物和骨板的接触点观察到。表面改性的生物医学钛合金具有更好的抗微动损伤的能力。事实证明,PVD TiN涂层和等离子氮化可有效减少摩擦并延迟材料的失效。在本研究中,已尝试通过摩擦测量和显微镜检查来解释PVD TiN涂层和等离子氮化Ti-6Al-4V和Ti-6Al-4V偶的微动疲劳破坏机理顺序。