首页> 外文期刊>Tierarztliche Praxis, Ausgabe K. Kleintiere >In vitro biomechanical comparison of a newly designed interlocking nail system to a standard DCP:Testing of cat femora in an osteotomy gap model
【24h】

In vitro biomechanical comparison of a newly designed interlocking nail system to a standard DCP:Testing of cat femora in an osteotomy gap model

机译:新设计的互锁钉系统与标准DCP的体外生物力学比较:截骨间隙模型中猫股骨的测试

获取原文
获取原文并翻译 | 示例
           

摘要

To describe a newly designed interlocking nail system (Targonr Vet System, TVS) tested in a model of diaphyseal femoral fractures in cats. Material and methods: Introduction of the TVS and presentation of the system components. Evaluation of application range and biomechanical testing of the TVS in cadaver bones under cyclic loading until fatigue failure occurred. The first two test groups compared the influence of implantation and immediate removal of the TVS locking bolts and six holes created by 2.0 mm cortical screws on the stability of feline femora. In the third group the two fixation systems were compared to each other with implants in place in an osteotomy gap model. The failure mode was statistically compared for each group (p < 0.05). Resu
机译:描述一种新设计的互锁式指甲系统(Targonr Vet System,TVS),该系统在猫的干phy股骨骨折模型中进行了测试。材料和方法:TVS的介绍和系统组件的介绍。评估尸体骨骼中TVS在循环载荷下的应用范围和生物力学测试,直到发生疲劳破坏为止。前两个测试组比较了植入和立即移除TVS锁定螺栓以及2.0 mm皮质螺钉产生的六个孔对猫股骨稳定性的影响。在第三组中,将两个固定系统在截骨间隙模型中放置有植入物的情况下相互比较。对每组的失败模式进行统计学比较(p <0.05)。 su

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号