Comparison of stent length reported by the stent's manufacturer to that determined by quantitative coronary angiography at the time of implantation versus that determined by coronary computed tomographic angiography at a later time

机译:支架制造商报告的支架长度与植入时通过定量冠状动脉造影确定的支架长度与以后使用冠状动脉计算机断层造影术确定的支架长度的比较

获取原文
获取原文并翻译 | 示例
           

摘要

The aim of this study was to assess whether coronary computed tomographic angiography (CCTA) would allow accurate length measurements that may help select optimal stent length. In 19 patients who underwent stent implantation and CCTA, the lengths of 30 implanted stents were assessed independently using quantitative coronary angiography (QCA) at the time of implantation and CCTA during follow-up. Measurements on CCTA and QCA were compared with the actual lengths of the implanted stents. The mean actual stent length was 17.3 ± 6.3 mm; the mean length of the stents measured using CCTA was 17.25 ± 5.54 mm, and the mean length of the stents measured using QCA was 15.92 ± 5.86 mm. There was a stronger positive correlation between measurements on CCTA versus actual stent lengths (R = 0.99, p = 0.0001) than between measurements on QCA versus actual stent lengths (R = 0.82, p = 0.0001) (p <0.0001 for the difference between correlation coefficients). In conclusion, the length measurements obtained with CCTA correlate better with an anatomic gold standard (actual stent lengths) than those obtained using QCA.
机译:这项研究的目的是评估冠状动脉计算机断层血管造影(CCTA)是否可以进行准确的长度测量,从而有助于选择最佳的支架长度。在19例接受了支架植入和CCTA的患者中,在植入时采用定量冠状动脉造影(QCA)并在随访期间进行了CCTA,分别评估了30个已植入支架的长度。将CCTA和QCA的测量值与植入支架的实际长度进行比较。平均实际支架长度为17.3±6.3 mm;使用CCTA测量的支架的平均长度为17.25±5.54 mm,使用QCA测量的支架的平均长度为15.92±5.86 mm。 CCTA测量值与实际支架长度之间的正相关性(R = 0.99,p = 0.0001)比QCA测量值与实际支架长度之间的正相关性(R = 0.82,p = 0.0001)(相关性之间的差异为p <0.0001)系数)。总之,与使用QCA获得的长度测量值相比,通过CCTA获得的长度测量值与解剖学金标准(实际支架长度)更好地相关。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号