Femoral sizer design can increase anterior notching during total knee arthroplasty

机译:股骨定尺设计可在全膝关节置换术中增加前切口

获取原文
获取原文并翻译 | 示例
           

摘要

Background: The anteroposterior (AP) portions of the distal femur are generally resected externally rotated relative to the posterior condylar line (PCL) in non-navigated/measured-resection total knee arthroplasty (TKA). Some femoral sizing devices allow rotational adjustment before sizing (rotation-modified sizing devices), while others do not (rotation-unmodified sizing devices). When rotation-unmodified sizing devices are used, the stylus attached to the sizing device identifies the proximal/anterior flange cut assuming the femoral component is aligned parallel to the PCL. However, if the AP cutting device is externally rotated then unexpected notching of the anterior cortex can occur.
机译:背景:股骨远端的前后(AP)部分通常在非导航/测量切除全膝关节置换术(TKA)中相对于后con突线(PCL)向外旋转。一些股骨上衣装置允许在上衣之前进行旋转调节(旋转修改过的上衣装置),而另一些则不允许(旋转不变的上衣装置)。当使用未旋转的定型设备时,假定股骨组件与PCL平行对齐,则连接到定型设备上的手写笔可识别近端/前法兰切口。但是,如果将AP切割装置从外部旋转,则可能会发生前皮质的意外切口。