Effect of 25(OH) vitamin D reference method procedure (RMP) alignment on clinical measurements obtained with the IDS-iSYS chemiluminescent-based automated analyzer

机译:25(OH)维生素D参考方法程序(RMP)对准对基于IDS-iSYS化学发光自动分析仪获得的临床测量结果的影响

获取原文
获取原文并翻译 | 示例

摘要

The Vitamin D Standardization Program (VDSP) has identified ID-LC/MS/MS as the reference method procedure (RMP) for 25(OH) vitamin D and NIST Standard SRM2972 as the standard reference material (SRM). As manufacturers align their products to the RMP and NIST standard, a concern is that results obtained in aligned assays will be divergent from those obtained with pre-alignment assays. The Immunodiagnostic Systems Ltd., chemiluminescent, 25(OH) vitamin D iSYS platform assay, was recently harmonized to the RMP. To determine the impact of standardization on results obtained with iSYS reagents, 119 single donor serum samples from eight different disease categories were analyzed in four non-standardized and two standardized iSYS assays. There were strong correlations between the four non-standardized and two standardized assays with Spearman's rank r values between 0.975 and 0.961 and four of the eight r values were >0.97. R-2 values for the eight best-fit linear regression equations ranging between 0.947 and 0.916. None of the slopes were found to be significantly different from one another. Bland-Altman plots showed that the bias was comparable when each of the four non-standardized assays was compared to either of the standardized assays. When the data were segregated in values between 6 and 49 ng/mL (15-122 nmol/L) or between 50 and 100 ng/mL (125-250 nmol/L) significant associations remained between results obtained with non-standardized and standardized calibrators regardless of the absolute value. When five recent DEQAS unknowns were analyzed in one non-standardized and one standardized assay the mean percent difference from the NIST target in values obtained using standardized vs. non-standardized calibrators were not significantly different. Finally, strong and statistically significant associations between the results were obtained using non-standardized and standardized assays for six of eight clinical conditions. The only exceptions were hypocalcemia and breast cancer, which likely reflect the small sample sizes for each of these diseases. These initial data provide confidence that the move to a NIST standardized assay will have little impact on results obtained with the iSYS platform.
机译:维生素D标准化计划(VDSP)已将ID-LC / MS / MS确定为25(OH)维生素D的参考方法程序(RMP),并将NIST标准SRM2972确定为标准参考材料(SRM)。当制造商将其产品调整为符合RMP和NIST标准时,需要关注的是,在比对测定中获得的结果与在比对前测定中获得的结果会有所不同。 Immunodiagnostic Systems Ltd.,化学发光的25(OH)维生素D iSYS平台测定法,最近与RMP协调使用。为了确定标准化对使用iSYS试剂获得的结果的影响,在四种非标准化和两种标准化的iSYS分析中分析了来自八种不同疾病类别的119个单一供体血清样品。 Spearman的秩r值在0.975和0.961之间,这四个非标准化测定和两个标准化测定之间存在很强的相关性,八个r值中的四个r> 0.97。八个最佳拟合线性回归方程的R-2值在0.947和0.916之间。没有发现坡度彼此之间有显着差异。布兰德-奥特曼(Bland-Altman)图显示,当将四种非标准化测定中的每一种与标准化测定中的任何一种进行比较时,偏差是可比较的。当数据以6到49 ng / mL(15-122 nmol / L)或50到100 ng / mL(125-250 nmol / L)的值分开时,非标准化和标准化结果之间仍存在显着关联校准器,与绝对值无关。在一项非标准化和一项标准化测定中分析了五个最近的DEQAS未知数时,使用标准化和非标准化校正剂获得的值与NIST目标的平均百分比差异无显着差异。最后,对于八种临床疾病中的六种,使用非标准化和标准化测定可以得到结果之间的强关联和统计学上的显着关联。低钙血症和乳腺癌是唯一的例外,这很可能反映出每种疾病的样本量都较小。这些初始数据提供了信心,即转向NIST标准化测定对使用iSYS平台获得的结果影响很小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号