STATEMENT OF PROBLEM: There are several electronic shade-matching instruments available for clinical use, but the reliability and accuracy of these instruments have not been thoroughly investigated. PURPOSE: The purpose of this in vitro study was to evaluate the reliability and accuracy of 4 dental shade-matching instruments in a standardized environment. MATERIAL AND METHODS: Four shade-matching devices were tested: SpectroShade, ShadeVision, VITA Easyshade, and ShadeScan. Color measurements were made of 3 commercial shade guides (Vitapan Classical, Vitapan 3D-Master, and Chromascop). Shade tabs were placed in the middle of a gingival matrix (Shofu GUMY) with shade tabs of the same nominal shade from additional shade guides placed on both sides. Measurements were made of the central region of the shade tab positioned inside a black box. For the reliability assessment, each shade tab from each of the 3 shade guide types was measured 10 times. For the accuracy assessment, each shade tab from 10 guides of each of the 3 types evaluated was measured once. Differences in reliability and accuracy were evaluated using the Standard Normal z test (2 sided) (alpha=.05) with Bonferroni correction. RESULTS: Reliability of devices was as follows: ShadeVision, 99.0%; SpectroShade, 96.9%; VITA Easyshade, 96.4%; and ShadeScan, 87.4%. A significant difference in reliability was found between ShadeVision and ShadeScan (P=.008). All other comparisons showed similar reliability. Accuracy of devices was as follows: VITA Easyshade, 92.6%; ShadeVision, 84.8%; SpectroShade, 80.2%; and ShadeScan, 66.8%. Significant differences in accuracy were found between all device pairs (P<.001) for all comparisons except for SpectroShade versus ShadeVision (P=.033). CONCLUSIONS: Most devices had similar high reliability (over 96%), indicating predictable shade values from repeated measurements. However, there was more variability in accuracy among devices (67-93%), and differences in accuracy were seen with most device comparisons.
机译:问题陈述:有几种可用于临床的电子阴影匹配仪器,但是尚未对这些仪器的可靠性和准确性进行彻底研究。目的:这项体外研究的目的是在标准化环境中评估4种牙科阴影匹配仪的可靠性和准确性。材料和方法:测试了四种阴影匹配设备:SpectroShade,ShadeVision,VITA Easyshade和ShadeScan。颜色测量是由3种商业阴影指南(Vitapan Classical,Vitapan 3D-Master和Chromascop)进行的。将遮光片放置在牙龈基质(Shofu GUMY)的中间,两侧的附加遮光板具有相同的标称阴影的遮光片。测量位于黑匣子内部的阴影选项卡的中央区域。为了进行可靠性评估,对3种阴影指南类型中的每种阴影标签进行了10次测量。为了进行准确性评估,对评估的3种类型的10种指南中的每个阴影选项卡进行了一次测量。使用带有Bonferroni校正的标准Normal z检验(2面)(alpha = .05)评估了可靠性和准确性上的差异。结果:设备的可靠性如下:ShadeVision,99.0%; SpectroShade,96.9%; VITA Easyshade,96.4%;和ShadeScan,占87.4%。在ShadeVision和ShadeScan之间发现了可靠性上的显着差异(P = .008)。所有其他比较显示出相似的可靠性。设备的准确性如下:VITA Easyshade,92.6%; ShadeVision,84.8%; SpectroShade,80.2%; ShadeScan占66.8%。对于所有比较而言,除SpectroShade与ShadeVision(P = .033)之外,所有器件对之间的准确性均存在显着差异(P <.001)。结论:大多数设备具有相似的高可靠性(超过96%),表明重复测量可预测阴影值。但是,设备之间的准确性差异更大(67-93%),大多数设备比较都发现准确性差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号