The reliability of radiographic assessment of femoral neck-shaft and implant angulation in hip resurfacing arthroplasty.

机译:髋关节表面置换术中股骨颈干和植入物成角的影像学评估的可靠性。

获取原文
获取原文并翻译 | 示例
           

摘要

Fifteen sets of patient radiographs were analyzed by 3 different observers on 2 occasions. Each observer measured the femoral neck-shaft angles (NSAs) of the preoperative digital radiographs and stem-shaft angles (SSAs) of the postoperative radiographs. The effect of femur position on SSA measured by digital radiographs was also investigated using a resurfaced synthetic femur. Radiographs were taken with the synthetic specimen positioned in 10 degrees increments of either flexion or rotation. Measurement by digital radiographs proved less than optimal in assessing preoperative NSA but was better in assessing the postoperative component SSA. External rotation of 30 degrees and flexion of 40 degrees resulted in a clinically significant disparity in SSA measurements. Patient malposition during radiographic imaging can contribute to erroneous NSA and SSA results.
机译:3位不同的观察者对2组进行了15组患者射线照相分析。每个观察者测量术前数字X线照片的股骨颈轴角(NSA)和术后X线照片的股骨干轴角(SSA)。还使用重铺表面的合成股骨研究了通过数字射线照相测量的股骨位置对SSA的影响。放射线照相是将合成样本放置在屈曲或旋转角度为10度的位置上。通过数字X射线照片测量证明在评估术前NSA方面并非最佳选择,但在评估术后SSA方面则更好。 30度的外部旋转和40度的弯曲导致SSA测量的临床显着差异。影像学检查期间患者错位可能导致错误的NSA和SSA结果。