Much evidence suggests that ultra high molecular weight polyethylene (UHMWP) irradiated to 9.5 Mrad has lower wear than UHMWP given 5 Mrad. Curiously, highly cross-linked polyethylenes gain weight during hip simulator testing. We postulated that (a) UHMWP irradiated to 9.5 Mrad would wear less than UHMWP irradiated to 5 Mrad UHMWP, (b) articulation against 38-mm heads would accentuate this difference, and (c) weight gain of highly cross-linked material reflects the inadequacy of load soak controls. We compared 9.5 versus 5 Mrad polyethylene in hip simulator wear tests, with both load soak corrections and with novel "motion soak" corrections. The 9.5-Mrad material wore less than 5-Mrad material for 32- and 38-mm heads. Motion soak corrections were more accurate than load soak corrections.
机译:许多证据表明,辐照至9.5 Mrad的超高分子量聚乙烯(UHMWP)的磨损低于5 Mrad的UHMWP。奇怪的是,在髋关节模拟器测试期间,高度交联的聚乙烯会增加重量。我们假设(a)辐照9.5 Mrad的UHMWP的磨损要少于辐照5 Mrad UHMWP的UHMWP,(b)38 mm头的铰接会加剧这种差异,并且(c)高交联材料的重量增加反映了负载保持控件的不足。我们在髋部模拟器磨损测试中比较了9.5 Mrad聚乙烯和5 Mrad聚乙烯,并进行了负载均湿校正和新型“运动均湿”校正。对于32毫米和38毫米磁头,9.5毫拉德材料的磨损小于5毫拉德材料。运动保持校正比负载保持校正更准确。