Oxidation in Ultrahigh Molecular Weight Polyethylene and Cross-Linked Polyethylene Acetabular Cups Tested against Roughened Femoral Heads in a Hip Joint Simulator

机译:在髋关节模拟器中针对粗大的股骨头测试超高分子量聚乙烯和交联聚乙烯髋臼杯中的氧化

获取原文
获取原文并翻译 | 示例
           

摘要

This study was aimed at comparing the oxidative degradation of commercial acetabular cups made of cross-linked polyethylene (XLPE) and conventional ultrahigh molecular weight polyethylene (UHMWPE).After testing against deliberately scratched CoCrMo femoral heads in a hip joint simulator,the cups,microtomed parallel to the articulating surface,were analyzed by IR spectroscopy.Due to the potential for artifacts caused by absorbed contaminants,the IR spectra were compared only after hexane extraction; actually,XLPE was found to absorb more serum than UHMWPE.The two sets of unworn acetabular cups showed different oxidation patterns with consequently different distributions of carbonyl species; unworn XLPE was characterized by lower contents of carbonyl species and hydrogen-bonded alcohols and higher contents of trans-vinylene species than unworn UHMWPE.Upon simulator testing,UHMWPE showed more significant changes in oxidation indexes and distribution of carbonyl compounds than XLPE,confirming a better wear behavior for XLPE under the adopted testing conditions.
机译:这项研究旨在比较由交联聚乙烯(XLPE)和常规超高分子量聚乙烯(UHMWPE)制成的商业髋臼杯的氧化降解。在髋关节模拟器中针对故意刮擦的CoCrMo股骨头进行测试后,将杯切成细小由于与被吸收的污染物可能产生假象,因此仅在萃取己烷后才比较IR光谱。实际上,发现XLPE比UHMWPE吸收更多的血清。两组未佩戴过的髋臼杯显示出不同的氧化模式,因此羰基物种的分布也不同。未磨损的XLPE的特征是与未磨损的UHMWPE相比,羰基物种和氢键醇的含量较低,反乙烯基乙烯的含量更高。在模拟器测试中,UHMWPE的氧化指数和羰基化合物分布的变化比XLPE更显着,证实了更好的性能XLPE在采用的测试条件下的磨损行为。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号