Minimum 10-Year Wear Analysis of Highly Cross-Linked Polyethylene in Cementless Total Hip Arthroplasty

机译:非骨水泥全髋关节置换术中高度交联聚乙烯的至少10年磨损分析

获取原文
获取原文并翻译 | 示例
           

摘要

Fifty-four patients (64 hips) underwent cementless total hip arthroplasty between 2000 and 2003 with a 22-mm zirconia ceramic bearing on highly cross-linked polyethylene, and were evaluated with a mean 11.9-year postoperative follow-up (range, 10-14 years). Linear wear was measured on the anteroposterior radiograph of the hip. No evidence of osteolysis and loosening was found on the final radiograph in any of the cases, and the steady-state linear wear rate was 0.017 +/- 0.018 mm/year. No significant correlation was found between the linear wear rate and age, body weight, cup inclination angle, or polyethylene thickness. Highly cross-linked polyethylene showed excellent wear resistance for >10 years when used in combination with 22-mm zirconia heads. (C) 2015 Elsevier Inc. All rights reserved.
机译:在2000年至2003年之间,对54例(64髋)患者进行了非骨水泥全髋关节置换术,并使用22毫米氧化锆陶瓷和高度交联的聚乙烯作为载体,并进行了平均11.9年的术后随访(范围为10- 14年)。在髋关节前后片上测量线性磨损。在任何情况下,在最终的X光片上均未发现骨溶解和松动的迹象,并且稳态线性磨损率为0.017 +/- 0.018毫米/年。在线性磨损率与年龄,体重,杯倾斜角或聚乙烯厚度之间没有发现显着相关性。当与22毫米氧化锆封头结合使用时,高度交联的聚乙烯在超过10年的时间内表现出优异的耐磨性。 (C)2015 Elsevier Inc.保留所有权利。