首页> 外文期刊>The Journal of arthroplasty >Corrosion Behavior of Tantalum-Coated Cobalt-Chromium Modular Necks Compared to Titanium Modular Necks in a Simulator Test
【24h】

Corrosion Behavior of Tantalum-Coated Cobalt-Chromium Modular Necks Compared to Titanium Modular Necks in a Simulator Test

机译:在模拟试验中,钽涂层钴铬模块化颈部与钛模块化颈部的腐蚀行为

获取原文
获取原文并翻译 | 示例
           

摘要

This study compared the corrosion behavior of tantalum-coated cobalt-chromium modular necks with that of titanium alloy modular necks at their junction to titanium-alloy femoral stem. Tests were performed in a dry assembly and two wet assemblies, one contaminated with calf serum and the other contaminated with calf serum and bone particles. Whereas the titanium modular neck tested in the dry assembly showed no signs of corrosion, the titanium modular necks tested in both wet assemblies showed marked depositions and corrosive attacks. By contrast, the tantalum-coated cobalt-chromium modular necks showed no traces of corrosion or chemical attack in any of the three assemblies. This study confirms the protective effect of tantalum coating the taper region of cobalt-chromium modular neck components, suggesting that the use of tantalum may reduce the risk of implant failure due to corrosion.
机译:这项研究比较了钽涂层钴铬组件颈部与钛合金组件颈部与钛合金股骨柄交界处的腐蚀行为。在干式装配和两个湿式装配中进行测试,一个被小牛血清污染,另一个被小牛血清和骨颗粒污染。在干燥组件中测试的钛模块颈部没有腐蚀迹象,而在两个湿组件中测试的钛模块颈部均显示出明显的沉积和腐蚀。相比之下,在三个组件中的任何一个组件上,涂有钽的钴铬模块颈部均未显示出腐蚀或化学侵蚀的痕迹。这项研究证实了钽对钴铬模块化颈部组件锥形区域的覆盖作用,表明钽的使用可以降低由于腐蚀造成的植入失败的风险。