The accuracy of computed tomography for determining femoral and tibial total knee arthroplasty component rotation.

机译:用于确定股骨和胫骨全膝关节置换术组件旋转的计算机断层扫描的准确性。

获取原文
获取原文并翻译 | 示例
       

摘要

Patellofemoral complications, instability, and tibial polyethylene wear after total knee arthroplasty (TKA) resulting from malrotation of the tibial or femoral components (or both) may be difficult to diagnose based on physical examination and standard knee radiographs. The preoperative assessment of implant rotational alignment is critical in planning treatment because the femoral or tibial component (or both) may need to be revised if malpositioned. The purpose of this study was to ascertain the accuracy of computed tomography (CT) scan for determining rotational alignment of femoral and tibial components in TKA. TKA components were inserted in human cadaver specimens at neutral and 5 degrees of external or internal rotation. For each position, the amount of rotation, determined from digital photographs, was compared with CT scan. The correlation coefficient between these two values averaged 0.87, which was significant at P < .05. The CT scan protocol described in this study can be applied clinically to patients with patellofemoral complaints to confirm or rule out the presence of component malrotation.
机译:基于体格检查和标准膝部X光片检查,可能难以诊断出因胫骨或股骨组件(或两者)变形而导致的全膝关节置换术(TKA)后的ello股并发症,不稳定性和胫骨聚乙烯磨损。植入物旋转对准的术前评估在计划治疗中至关重要,因为如果放置不当,可能需要修改股骨或胫骨组件(或两者)。这项研究的目的是确定计算机断层扫描(CT)扫描确定TKA中股骨和胫骨组件旋转对齐的准确性。将TKA组件以中性和5度外部或内部旋转角度插入人体尸体标本中。对于每个位置,将从数字照片确定的旋转量与CT扫描进行比较。这两个值之间的相关系数平均为0.87,在P <.05时很显着。这项研究中描述的CT扫描方案可以临床应用于pa股主诉的患者,以确认或排除存在组件错误的情况。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号