首页> 外文期刊>The Journal of arthroplasty >Polyethylene wear against alumina and zirconia heads in cemented total hip arthroplasty.
【24h】

Polyethylene wear against alumina and zirconia heads in cemented total hip arthroplasty.

机译:在固定全髋关节置换术中,聚乙烯会磨损氧化铝和氧化锆头。

获取原文
获取原文并翻译 | 示例

摘要

We compared the polyethylene wear of acetabular sockets articulated with 22.225-mm alumina heads with the polyethylene wear of those articulated with 22.225-mm zirconia heads in cemented total hip arthroplasty during a mean follow-up period of 5.4 years. Using a computer-aided technique, we measured polyethylene wear radiologically in 46 hips with alumina heads and 58 hips with zirconia heads. The preoperative diagnosis in all cases was osteoarthritis. The mean linear wear rate and mean volumetric wear rate of polyethylene sockets against zirconia heads were 0.133 mm/y and 39.8 mm(3)/y, respectively, significantly greater (P < .01) than the wear rates against alumina heads (0.078 mm/y and 24.2 mm(3)/y, respectively). Age at operation, patient body weight as well as height, thickness of polyethylene, and socket abduction angle did not influence the wear rates. We speculate that the excessive polyethylene wear was caused by phase transformation of zirconia, leading to an increase of surface roughness.
机译:我们比较了在平均5.4年的随访期内,在全髋关节置换术中,用22.225-mm氧化铝头连接的髋臼窝的聚乙烯磨损与用22.225-mm氧化锆头连接的髋臼的聚乙烯磨损。使用计算机辅助技术,我们用放射学方法测量了46头氧化铝头和58头氧化锆头的聚乙烯的聚乙烯磨损。所有病例的术前诊断均为骨关节炎。聚乙烯套头对氧化锆头的平均线性磨损率和平均体积磨损率分别为0.133 mm / y和39.8 mm(3)/ y,比氧化铝头的磨损率(0.078 mm)大得多(P <.01) / y和24.2 mm(3)/ y)。手术年龄,患者体重以及身高,聚乙烯厚度和承窝外展角度均不会影响磨损率。我们推测聚乙烯的过度磨损是由于氧化锆的相变引起的,从而导致表面粗糙度的增加。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号