Wear rates of ceramic-on-ceramic bearing surfaces in total hip implants: a 12-year follow-up study.

机译:全髋关节植入物中陶瓷对陶瓷的支承面的磨损率:一项为期12年的随访研究。

获取原文
获取原文并翻译 | 示例
           

摘要

A retrospective clinical and radiographic analysis was performed on 58 patients (60 hips; mean age at time of surgery, 45.2 years) at a minimum of 10-year follow-up (mean, 12.7 years) after total hip replacement using a ceramic-on-ceramic hearing total hip implant (Autophor, Smith and Nephew, Memphis, TN). Mean wear rate at final follow-up was 0.21 mim, averaging 0.016 mm/y. There were no cases of periprosthetic osteolysis in the acetabuulum or femur. For the unrevised components, there were 3 (5%) cases of protrusio acetabuli and 4 (7%) cases of acetabular component loosening. On the femoral side, 78.3% had distal pedestal formation, and 83% had greater than 2 mm implant-bone radiolucencies in more than 5 Gruen zones as a result of gross motion of the stem. Despite radiographic evidence of implant loosening, this hard bearing articulation functioned well in vivo for more than 12 years with remarkably low wear--approximately one tenth the rate reported for metal-on-polyethylene total hip bearings.
机译:对58例患者(60髋;手术时的平均年龄,45.2岁)进行了回顾性临床和放射学分析,至少在陶瓷置换全髋置换术后10年(平均12.7年)进行了随访。 -陶瓷听力全髋关节植入物(田纳西州孟菲斯的Autophor,Smith和Nephew)。最终随访时的平均磨损率为0.21 mim,平均为0.016 mm / y。在髋臼或股骨中没有假体周围骨溶解的病例。对于未修订的组件,有3例(5%)髋臼隆突和4例(7%)髋臼假体松动。在股骨侧,由于茎干运动,在超过5个Gruen区域中,有78.3%的患者具有远端基座形成,而83%的患者的植入骨射线透透性大于2 mm。尽管有放射学证据表明植入物松动,但这种坚硬的关节关节在体内功能良好,使用寿命超过12年,磨损极低,大约是金属对聚乙烯全髋关节轴承报道率的十分之一。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号