Differences in the rates of penetration determined from radiographic and shadowgraphic measurements of acetabular sockets.

机译:穿透率的差异由髋臼窝的射线照相和阴影照相测量确定。

获取原文
获取原文并翻译 | 示例
       

摘要

A number of previous studies have observed a marked difference in the penetration depths recorded between direct and radiographic measurement, although no assessment of how these discrepancies affect the rates of penetration has been undertaken. In this report, the penetration depth of 95 sockets was determined from both prerevision radiographs and casts of the retrieved sockets using the uniradiographic and shadowgraphic techniques, respectively. It was observed that the mean discrepancy between the penetration rates derived from the two measurement methods was 0.046 (SE 0.017) mm yr(-1), which was significantly different from zero at P = .007. It was concluded that the penetration rates derived from this method of radiographic assessment seriously underestimated the true value as measured by a direct method by approximately 20%. As a consequence, the discrepancy in penetration rates between those sockets that are loose at revision surgery (direct measurement) and the general population of acetabular components (radiographic measurement) may not be as large as previously thought.
机译:尽管尚未评估这些差异如何影响穿透率,但许多先前的研究已经观察到直接和射线照相测量之间记录的穿透深度有显着差异。在该报告中,分别使用单射线照相技术和阴影成像技术从预修订射线照相和取回的套模铸件中确定了95个套环的穿透深度。观察到,两种测量方法得出的穿透率之间的平均差异为0.046(SE 0.017)mm yr(-1),在P = .007时与零有显着差异。结论是,从这种射线照相评估方法得出的穿透率严重低估了通过直接方法测得的真实值约20%。结果,在翻修手术(直接测量)时松动的那些窝与髋臼假体的总体数量(放射线照相)之间的穿透率差异可能不会像以前想象的那么大。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号