We compared the survival outcomes, left ventricular assist device (LVAD)-related hospitalization, stroke, infection, panel reactive antibody, and blood product use data among 13 Novacor and 51 HeartMate system recipients. Stroke was significantly higher in Novacor patients, as was blood product use at the time of heart transplantation, likely due to long-term anti-coagulation, while the LVAD-related hospitalization and infections did not differ between the 2 groups. A positive panel reactive antibody was seen more among the HeartMate patients, but did not have a significant clinical impact and may not represent a true allosensitization.
机译:我们比较了13位Novacor和51位HeartMate系统接受者的生存结局,左心室辅助设备(LVAD)相关的住院治疗,中风,感染,面板反应性抗体和血液制品使用数据。 Novacor患者的卒中显着升高,心脏移植时使用血液制品的卒中显着更高,这可能是由于长期的抗凝治疗所致,而LVAD相关的住院和感染在两组之间没有差异。在HeartMate患者中,阳性反应性抗体阳性率更高,但对临床无显着影响,可能并不代表真正的同种过敏。