A comparative study of mini-external fixation systems used to treat unstable metacarpal fractures.

机译:微型外部固定系统用于治疗不稳定的掌骨骨折的比较研究。

获取原文
获取原文并翻译 | 示例
           

摘要

We compared the biomechanical properties, clinical versatility, application ease, and financial cost of 7 mini-external fixation systems used to treat unstable metacarpal shaft fractures with segmental bone loss. Attached to a maple dowel fracture model, each fixator was tested in axial compression, lateral 4-point bending, 45 degrees apex-dorsal 4-point bending, and torsion. The EBI DynaFix system had the highest overall rigidity values and was highly versatile; the Synthes/AO Mini-External Fixator had intermediate rigidity values and was the lightest system tested; the Howmedica Mini-Hoffmann fixator was substantially less rigid than all the other tested constructs but was the most versatile. Devices constructed with polymethylmethacrylate and Kirschner wires were rigid, lightweight, and inexpensive. Routine use of these devices is discouraged, however, because of unacceptable loosening at the pin-cement interface during testing and because of difficulties encountered during construction and application. None of the external fixation systems tested was superior in all the parameters analyzed.
机译:我们比较了用于治疗不稳定的掌骨干骨折伴节段性骨丢失的7种微型外固定系统的生物力学性能,临床通用性,易用性和财务成本。固定在枫木钉断裂模型上,对每个固定器进行轴向压缩,横向4点弯曲,45度顶背4点弯曲和扭转测试。 EBI DynaFix系统具有最高的整体刚度值,并且用途广泛。 Synthes / AO微型外部固定器具有中等的刚性值,并且是最轻的测试系统; Howmedica Mini-Hoffmann固定器的刚性明显低于所有其他测试结构,但用途最多。用聚甲基丙烯酸甲酯和克氏针制成的设备坚固,轻巧且便宜。但是,由于在测试过程中销钉-水泥界面处的松动是不可接受的,并且在构造和应用过程中遇到了困难,因此不建议常规使用这些设备。在所有分析的参数中,所测试的外部固定系统均无一例外。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号