A voltammetric enzyme-linked immunoassay based on new system of m-aminophenol (MAP)-H_2O_2-horseradish peroxidase (HRP) has firstly been developed and used for the detection of HPR, labelled HRP and ferritin in human serum. HRP or labelled HRP catalyzes the oxidation reaction of MAP with H_2O_2, the product of which produces a sensitive voltammetric peak at potential of -0.46 V (vs. SCE) in Britton-Robinson (BR) buffer solution. By using this voltammetric peak, HRP can be measured with a detection limit of 9.5 * 10~(-1) mU/1 and a linear range of 2.5 - 2.5 * 10~2 mU/1. The detection limit of ferritin is 0.25 ng/ml and the linear range 0.25-320 ng/ml. The processes of the electro-reduction of the product of the enzyme-catalyzed reaction have been investigated in detail.
机译:首次开发了一种基于间氨基苯酚(MAP)-H_2O_2-辣根过氧化物酶(HRP)新系统的伏安酶联免疫法,并将其用于检测人血清中的HPR,标记的HRP和铁蛋白。 HRP或标记的HRP催化MAP与H_2O_2的氧化反应,在Britton-Robinson(BR)缓冲溶液中,H_2O_2的产物在-0.46 V(相对于SCE)的电势下产生灵敏的伏安峰。通过使用该伏安峰,可以检测到HRP的检出限为9.5 * 10〜(-1)mU / 1,线性范围为2.5-2.5 * 10〜2 mU / 1。铁蛋白的检出限为0.25 ng / ml,线性范围为0.25-320 ng / ml。已经详细研究了酶催化反应产物的电还原过程。