Influence of Measurement Depth on the Stiffness Assessment ofHealthy Liver with Real-Time Shear Wave Elastography

机译:实时剪切波弹性成像技术对测量深度对健康肝脏硬度评估的影响

获取原文
获取原文并翻译 | 示例
       

摘要

The purpose of this study was to determine the measurement depth range within which liver stiffness can be reliably assessed using real-time shear wave elastography (SWE) technology. Measurements were performed on phantoms and healthy volunteers. In the first group of patients, measurements were performed at depths of 2-8cm from the probe surface. In the second group of patients, measurements were conducted 0-7cm below the liver capsule. Success rate of measurements (SRoM), success rate of patients (SRoS) and coefficients of variation (CVs) of repeated measurements were compared. The SRoMs at 3-7cm and the CVs at 2-5cm from the probe surface were significantly higher and lower than those at other depths (p<0.001), respectively. SRoS was zero 0-1cm below the liver capsule. Furthermore, the features of 2-D stiffness mapping images were also found to change with depth. According to our results, the depth range for the most reliable liver stiffness assessment using SWE should be 3-5cm from the probe surface and simultaneously 1-2cm below the liver capsule.
机译:这项研究的目的是确定测量深度范围,在该范围内可以使用实时剪切波弹性成像(SWE)技术可靠地评估肝硬度。在幻像和健康志愿者身上进行测量。在第一组患者中,在距探头表面2-8cm的深度进行测量。在第二组患者中,在肝囊下方0-7厘米处进行测量。比较了测量的成功率(SRoM),患者的成功率(SRoS)和重复测量的变异系数(CV)。距探头表面3-7cm处的SRoM和2-5cm处的CV分别显着高于和低于其他深度处的SRoM(p <0.001)。 SRoS在肝囊下方为0 0-1cm。此外,还发现二维刚度映射图像的特征随深度而变化。根据我们的结果,使用SWE进行最可靠的肝硬度评估的深度范围应距探头表面3-5cm,同时距肝囊下面1-2cm。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号