首页> 外文期刊>Ultrasonics sonochemistry >Towards a reference ultrasonic cavitation vessel. Part 1: Preliminary investigation of the acoustic field distribution in a 25 kHz cylindrical cell
【24h】

Towards a reference ultrasonic cavitation vessel. Part 1: Preliminary investigation of the acoustic field distribution in a 25 kHz cylindrical cell

机译:朝向参考超声空化容器。第1部分:25 kHz圆柱单元中声场分布的初步研究

获取原文
获取原文并翻译 | 示例

摘要

The acoustic field produced by a 25 kHz, 25 1 cylindrical sonochemical processing cell has been characterised systematically using a sonar hydrophone, with the aim of establishing it as a reference test bed on which future investigations into acoustic cavitation activity may be based. Data acquired at sonication levels up to 500 W have shown that though significant cavitation activity is generated throughout the vessel, the acoustic field generated is reproducible, typically to +/- 12%. The increases in acoustic pressure are shown to be nonlinear with applied power, suggesting an intermediate optimum level for future study. (c) 2006 Elsevier B.V. All rights reserved.
机译:已使用声纳水听器对25 kHz,25 1圆柱形声化学处理池产生的声场进行了系统表征,目的是将其建立为参考测试台,以作为将来对声空化活动进行研究的基础。在高达500 W的超声处理水平下获得的数据表明,尽管在整个血管中都产生了明显的空化活动,但所产生的声场是可重现的,通常为+/- 12%。声压的增加与施加的功率呈非线性关系,为将来的研究提供了一个中等的最佳水平。 (c)2006 Elsevier B.V.保留所有权利。