Fluid composition impacts standardized testing protocols in ultrahigh molecular weight polyethylene knee wear testing

机译:流体成分影响超高分子量聚乙烯膝盖磨损测试中的标准化测试方案

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Wear of total knee replacements is determined gravimetrically in simulator studies. A mix of bovine serum, distilled water, and additives is intended to replicate the lubrication conditions in vivo. Weight gain due to fluid absorption during testing is corrected using a load soak station. In this study, three sets of ultrahigh molecular weight polyethylene tibial plateau were tested against highly polished titanium condyles. Test 1 was performed in two different institutions on the same simulator according to the standard ISO 14243-1, using two testing lubricants. Test 2 and test 3 repeated both previous test sections. The wear and load soak rates changed significantly with the lubricant. The wear rate decreased from 16.9 to 7.9 mg weight loss per million cycles when switching from fluid A to fluid B. The weight gain of the load soak specimen submersed in fluid A was 6.1 mg after 5×10{sup}6 cycles, compared with 31.6 mg for the implant in fluid B after the same time period. Both lubricants were mixed in accordance with ISO 14243 (Implants for surgery - wear of total knee-joint prostheses), suggesting that calf serum should be diluted to 25 ± 2 per cent with deionized water and a protein mass concentration of not less than 17 g/l. The main differences were the type and amount of additives that chemically stabilize the lubricant throughout the test. The results suggest that wear rates can only be compared if exactly the same testing conditions are applied. An agreement on detailed lubricant specifications is desirable.
机译:在模拟器研究中通过重量分析确定全膝关节置换的磨损。牛血清,蒸馏水和添加剂的混合物旨在复制体内润滑条件。使用负载均压站校正了在测试过程中由于液体吸收而导致的重量增加。在这项研究中,针对高度抛光的钛con测试了三组超高分子量聚乙烯胫骨平台。根据ISO 14243-1标准,在1个模拟器上的两个不同机构中使用两种测试润滑剂进行了测试1。测试2和测试3重复了前面的两个测试部分。润滑剂的磨损率和负荷保持率明显变化。当从流体A转换为流体B时,磨损率从每百万个循环16.9重量损失减少到7.9毫克。浸没在流体A中的负载浸泡样本的重量增加是在5×10 6循环后的6.1毫克。在相同时间段后,对于B液中的植入物为31.6 mg。两种润滑剂均按照ISO 14243(外科手术的模拟物-整个膝关节假体的磨损)进行混合,建议用去离子水将小牛血清稀释至25±2%,蛋白质质量浓度应不少于17 g / l。主要区别在于在整个测试过程中化学稳定润滑剂的添加剂的类型和数量。结果表明,只有在完全相同的测试条件下才能比较磨损率。希望就详细的润滑剂规格达成一致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号