Improved procedures for the laboratory study of high-temperature viscoelastic relaxation

机译:改进的高温粘弹性松弛实验室研究程序

获取原文
获取原文并翻译 | 示例

摘要

Several aspects of the experimental procedure used in our laboratory for the study of high-temperature viscoelastic relaxation in geological and ceramic materials have been critically examined and refined. Most importantly, we have newly quantified a contribution to the torsional compliance of the specimen assembly from the interfaces between the metal foil and alumina torsion rod at each end of the specimen. We demonstrate that this extraneous contribution to the apparent compliance of the specimen can be removed by subtraction of the torsional compliance of a foil-bearing reference assembly similarly containing two alumina-foil interfaces. This new strategy along with other improvements including the modelling of drift in transducer sensitivity, suppression of background creep, minimising the aliasing of high-frequency noise, and allowance for the viscoelastic behaviour of the alumina control specimen, promise more accurate determination of the dispersion and attenuation associated with pronounced high-temperature viscoelastic relaxation. Correction of previously published data for fine-to-medium grained polycrystalline olivine for the interfacial compliance suggests somewhat milder frequency and temperature dependence of Q(-1) than previously reported and substantially stronger grain size sensitivity.
机译:我们的实验室用于研究地质和陶瓷材料中高温粘弹性松弛的实验程序的几个方面都经过了严格的检查和完善。最重要的是,我们从样品两端的金属箔和氧化铝扭力杆之间的界面上,重新量化了对样品组件扭转柔度的贡献。我们证明,通过减去类似地包含两个氧化铝-箔界面的带有箔的参考组件的扭转柔度,可以消除对样品的表面柔度的这种外来贡献。这一新策略以及其他改进措施包括对传感器灵敏度的漂移进行建模,抑制背景蠕变,最大程度地减少高频噪声的混叠以及对氧化铝对照样品的粘弹性行为的考虑,有望更准确地确定色散和与明显的高温粘弹性松弛相关的衰减。校正以前发布的中至中等粒度的多晶橄榄石的界面顺应性数据表明,Q(-1)的频率和温度依赖性比以前报道的要温和一些,并且晶粒尺寸的敏感性要强得多。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号