Background: Surgical site infection remains a concern in orthopedic surgery, and contamination of C-arm covers is a potentially modifiable risk factor. Methods: A single-cohort study was conducted using 30 consecutive patients undergoing operative fracture fixation. Cultures were obtained from the C-arm cover after initial draping and every 20 min thereafter. The total number of persons in the operating room (person-hours/h of study time) and the number of door openings were recorded. The C-arm position changes and the time to contamination were monitored. Results: The median time from the start of the operation to contamination was 20 min. There was a 17% contamination rate on initial draping, 50% at 20 min, 57% at 40 min, and 80% by 80 min. The C-arms in five cases were not contaminated during the surgery. Time to contamination correlated significantly with lateral position changes (correlation [r] = 0.64; p = 0.003) but was not related to C-arm position changes (r = 0.22; p = 0.34), number of door openings (r = 0.20; p = 0.39), or person-hours/h (r = 0.04; p = 0.85). Conclusions: Contamination of the C-arm drape occurs often and early during surgery for orthopedic fractures. We recommend minimal contact with the C-arm to avoid contamination of the surgical field.
机译:背景:手术部位感染仍然是整形外科中的一个问题,C型臂套的污染是潜在的可改变的危险因素。方法:对30例接受手术骨折固定的患者进行了单队列研究。最初悬垂后,然后每20分钟从C型臂盖中获得培养物。记录手术室中的总人数(人时/小时的学习时间)和开门次数。监控C型臂的位置变化和污染时间。结果:从手术开始到污染的中位时间为20分钟。初始悬垂时的污染率为17%,20分钟时为50%,40分钟时为57%,80分钟时为80%。手术中5例C型臂未受到污染。污染时间与侧向位置变化显着相关(相关度[r] = 0.64; p = 0.003),但与C型臂位置变化(r = 0.22; p = 0.34),门开数量(r = 0.20; p = 0.34)无关。 p = 0.39)或人/小时/小时(r = 0.04; p = 0.85)。结论:矫形骨折的手术中C形悬垂巾的污染经常发生,并且在手术早期。我们建议尽量避免与C型臂接触,以免污染手术区域。

著录项

  • 来源
    《Surgical infections》 |2012年第2期|p.110-113|共4页
  • 作者单位

    Department of Orthopedics, Wright State University Boonshoft School of Medicine, Dayton, Ohio;

    Department of Orthopedics, Wright State University Boonshoft School of Medicine, Dayton, Ohio;

    Department of Orthopedics Miami Valley Hospital Weber Building, 2nd Floor 128 E. Apple St. Dayton, OH 45409;

    Department of Orthopedics, Wright State University Boonshoft School of Medicine, Dayton, Ohio;

    Department of Orthopedics, Wright State University Boonshoft School of Medicine, Dayton, Ohio;

    Department of Orthopedics, Wright State University Boonshoft School of Medicine, Dayton, Ohio;

  • 收录信息 美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号