首页> 外文期刊>Sensors and Actuators >Detection of ochratoxin A in wine and beer by chemiluminescence-based ELISA in microfluidics with integrated photodiodes
【24h】

Detection of ochratoxin A in wine and beer by chemiluminescence-based ELISA in microfluidics with integrated photodiodes

机译:基于化学发光的ELISA结合光电二极管检测微流体中葡萄酒和啤酒中的och曲霉毒素A

获取原文
获取原文并翻译 | 示例
           

摘要

Ochratoxin A (OTA) is a mycotoxin produced by naturally occurring fungi in crops such as grapes and cereals. OTA has demonstrated toxicity and carcinogenicity in animals and therefore biosensing systems are required for its detection and quantification in commercialized goods. This work demonstrates an integrated analytical system that conjugates an indirect competitive enzyme-linked immunosorbent assay strategy developed in PDMS microfluidics with integrated microfabricated hydrogenated amorphous silicon photodiodes for chemiluminescence detection. A limit of detection of 0.85 ng/mL was obtained for OTA detection in a PBS solution using a straight-channel configuration. Comparable limits of detection were obtained for beer extracts but for red wine extracts a higher limit of detection of OTA of 28 ng/mL was obtained. A two-channel U-shaped microfluidic device was developed to perform the simultaneous analysis of a reference solution and of an ochratoxin A contaminated solution. This configuration efficiently reduced measurement errors resulting in an improvement of one order of magnitude in the limits of detection of OTA in beer and red wine extracts. The results show that the miniaturized system can be developed into one stage of a future highly sensitive, portable, and fully integrated "toxin-chip" for monitoring food safety applications.
机译:ch曲霉毒素A(OTA)是一种真菌毒素,是由葡萄和谷物等农作物中的天然真菌产生的。 OTA已证明对动物具有毒性和致癌性,因此在商业商品中对其进行检测和定量需要生物传感系统。这项工作演示了一个集成的分析系统,该系统结合了PDMS微流体技术中开发的间接竞争性酶联免疫吸附测定策略和用于化学发光检测的集成的微制造的氢化非晶硅光电二极管。对于使用直通道配置的PBS溶液中的OTA检测,检测限为0.85 ng / mL。啤酒提取物的检测限相当,但红酒提取物的OTA检出限更高,为28 ng / mL。开发了一种两通道U形微流控设备,可同时分析参考溶液和and曲霉毒素A污染溶液。这种配置有效地减少了测量误差,从而使啤酒和红酒提取物中OTA的检出限提高了一个数量级。结果表明,小型化系统可以发展成为未来高度敏感,便携式且完全集成的“毒素芯片”的一个阶段,以监控食品安全应用。

著录项

  • 来源
    《Sensors and Actuators》 |2013年第1期|232-240|共9页
  • 作者单位

    INESC Microsistemas e Nanotecnologias and IN-Institute ofNanoscience and Nanotechnology, Lisbon, Portugal;

    INESC Microsistemas e Nanotecnologias and IN-Institute ofNanoscience and Nanotechnology, Lisbon, Portugal;

    Department of Bioengineering, Instituto Superior Tecnico, Technical University of Lisbon, Lisbon, Portugal,IBB - Institute for Biotechnology and Bioengineering, Centre for Biological and Chemical Engineering, Instituto Superior Tecnico, Technical University of Lisbon, Lisbon, Portugal;

    INESC Microsistemas e Nanotecnologias and IN-Institute ofNanoscience and Nanotechnology, Lisbon, Portugal;

    INESC Microsistemas e Nanotecnologias and IN-Institute ofNanoscience and Nanotechnology, Lisbon, Portugal,Department of Bioengineering, Instituto Superior Tecnico, Technical University of Lisbon, Lisbon, Portugal;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    ochratoxin A; chemiluminescence ELISA; microfluidics; microfabricated photodiodes;