首页> 外文期刊>Sensors and Actuators >Rapid and multiple detections of staphylococcal enterotoxins by two-dimensional molecularly imprinted film-coated QCM sensor
【24h】

Rapid and multiple detections of staphylococcal enterotoxins by two-dimensional molecularly imprinted film-coated QCM sensor

机译:二维分子印迹膜包裹QCM传感器快速,多次检测葡萄球菌肠毒素

获取原文
获取原文并翻译 | 示例
           

摘要

A quartz crystal microbalance with dissipation biosensor based on two-dimensional (2D) molecularly imprinted film-coated by organosilanes and the template protein staphylococcal enterotoxins (SEs) in situ synthesizing on the surface of the piezoelectric quartz crystal (PQC) as transducer was developed for the detection of SEs in real samples. By incorporated with the sensor, frequency shifts (△fs) and dissipation shifts (△Ds) were recorded and monitored in real time for the evaluation of the property of the sol-gel imprinted films for the detection of targets. Both of the working ranges for the detection of SEA and SEB were 0.1 -1000 μg/mL with the lowest detection limits 7.97 and 2.25 ng/mL, respectively. Performances and surface structures of the fabricated 2D sol-gel films were also observed and confirmed by atomic force microscopy. Results demonstrated that the 2D molecularly imprinted film-coated QCM could give repeated usage and selective recognition to the template SEs. The intelligentized biosensor was employed to SEA and SEB detection in milk samples after pre-treatment. The recoveries ranged from 97.00% to 114.20% and it can be used for ten times indicating that the developed method was feasible and applicable in the detection of SEs.
机译:开发了一种基于耗散生物传感器的石英晶体微天平,该传感器基于二维(2D)分子印迹薄膜,该薄膜由有机硅烷和模板蛋白葡萄球菌肠毒素(SEs)包覆,作为传感器直接在压电石英晶体(PQC)的表面上合成。真实样品中SE的检测。通过与传感器结合,可以实时记录和监测频移(△fs)和耗散频移(△Ds),以评估溶胶凝胶印迹膜的特性,以检测目标。 SEA和SEB的两个检测范围均为0.1 -1000μg/ mL,最低检测限分别为7.97和2.25 ng / mL。还观察了所制造的2D溶胶-凝胶膜的性能和表面结构,并通过原子力显微镜确认。结果表明,二维分子印迹薄膜包被的QCM可以使模板SE重复使用和选择性识别。预处理后,智能化的生物传感器用于牛奶样品中的SEA和SEB检测。回收率从97.00%到114.20%不等,可重复使用10次,表明该方法可行,可用于SEs的检测。

著录项

  • 来源
    《Sensors and Actuators》 |2014年第2期|326-331|共6页
  • 作者单位

    Tianjin Key Laboratory of Risk Assessment and Control Technology for Environment and Food Safety, Institute of Health and Environmental Medicine,Tianjin 300050, PR China;

    Tianjin Key Laboratory of Risk Assessment and Control Technology for Environment and Food Safety, Institute of Health and Environmental Medicine,Tianjin 300050, PR China;

    Tianjin Key Laboratory of Risk Assessment and Control Technology for Environment and Food Safety, Institute of Health and Environmental Medicine,Tianjin 300050, PR China;

    Tianjin Key Laboratory of Risk Assessment and Control Technology for Environment and Food Safety, Institute of Health and Environmental Medicine,Tianjin 300050, PR China;

    Tianjin Key Laboratory of Risk Assessment and Control Technology for Environment and Food Safety, Institute of Health and Environmental Medicine,Tianjin 300050, PR China;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Staphylococcal enterotoxin; Two-dimensional; Quartz crystal microbalance with; dissipation; Molecularly imprinted polymer; Sol-gel; Lowest detection limit;

    机译:葡萄球菌肠毒素;二维;石英晶体微量天平;耗散;分子印迹聚合物;溶胶凝胶最低检测限;