The behavior of the same ion-selective electrode (ISE) under three measurement modes: potentiometric, chronoamperometric and coulometric is compared using valinomycin-based K~+-ISE as model system. The ISE response to decrease and to increase of the K~+ ion activity is studied systematically. It is shown that the theoretical description of the time dependencies of current and of charge must take the concentration polarization in consideration, and the respective equations are presented. The concentration of K~+ ion in blood serum samples is measured in the potentiometric, in the chronoamperometric and in the coulometric mode, and satisfactory results are obtained in all three cases.
机译:使用基于缬氨霉素的K〜+ -ISE作为模型系统,比较了同一离子选择电极(ISE)在三种测量模式下的行为:电位测量,计时电流测量和库仑测量。系统地研究了ISE对K〜+离子活性降低和增加的响应。结果表明,电流和电荷时间相关性的理论描述必须考虑浓度极化,并给出了相应的方程。用电位法,计时电流法和库仑法测量血清样品中的钾离子浓度,在这三种情况下均获得满意的结果。