摘要

We report spectrophotometric methods for quantifying sodium and potassium in serum and plasma without sample pretreatment or solvent-extraction steps. The methods are based on novel chromogenic ionophores characterized by molecular structures highly preorganized for binding, which result in exceptionally high sensitivity and selectivity (negligible responses to sodium at 80 to 200 mmol/L in the potassium assay and to potassium at 0 to 10 mmol/L in the sodium assay). The within-run imprecisions are comparable with those for ion-selective electrodes (for K, 1.0% at 4.5 mmol/L; for Na, 1.1% at 136 mmol/L); the analytical range is 2-10 mmol/L for potassium, 80-170 mmol/L for sodium. The results for greater than 100 patients' samples correlated well with results obtained with ion-selective electrodes (r = 0.99 for K, 0.97 for Na). The ready-to-use aqueous reagents will be available under the trade name of ChromoLyte for use with Technicon's RA family of clinical analyzers.
机译:我们报告了无需样品预处理或溶剂萃取步骤即可定量测定血清和血浆中钠和钾的分光光度法。该方法基于新颖的生色离子载体,其特征是高度预先结合的分子结构,从而具有极高的灵敏度和选择性(在钾测定中对80至200 mmol / L的钠和对0至10 mmol / L的钾的响应可忽略不计) L在钠含量测定中)。行程内的不精确度与离子选择电极相当(对于K,在4.5 mmol / L时为1.0%;对于Na,在136 mmol / L时为1.1%);对于离子选择性电极钾的分析范围是2-10 mmol / L,钠的分析范围是80-170 mmol / L。超过100个患者样品的结果与离子选择电极获得的结果非常相关(K = r = 0.99,Na = 0.97)。即用型水性试剂将以ChromoLyte的商品名提供,可与Technicon的RA临床分析仪系列一起使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号