MONTE CARLO METHOD FOR RADIOLOGICAL X-RAY EXAMINATIONS

机译:放射线X射线检查的蒙特卡洛方法

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The organ doses and the effective dose of patients exposed at an X-ray beam, having photon energies between 10-150 keV, are analyzed by using a new method named IradMed. Three major radiological procedures, mammography, radiography and Computer Tomography are considered. The Rayleigh scattering, the Compton scattering and the photoelectric effect were taken into account for patient dose calculations. The Compton scattering is modeled by using "classic", Kahn. Wielopolski and EGS procedures. The best results for angular distribution was obtained when the EGS-based algorithm was used. It can be noted that patient doses determined by using Kahn or "classic" procedures are in good agreement with data from literature when similar computing methods are used. From our data, for patient dose calculations, the EGS procedure is recommended.
机译:通过使用一种名为IradMed的新方法,分析了X射线束下光子能量在10-150 keV之间的患者的器官剂量和有效剂量。考虑了三个主要的放射学程序,乳腺摄影,放射摄影和计算机断层摄影。计算患者剂量时应考虑瑞利散射,康普顿散射和光电效应。康普顿散射是通过使用“经典”(Kahn)建模的。 Wielopolski和EGS程序。当使用基于EGS的算法时,可以获得最佳的角度分布结果。可以注意到,当使用类似的计算方法时,通过使用卡恩或“经典”程序确定的患者剂量与来自文献的数据高度吻合。根据我们的数据,为了进行患者剂量计算,建议使用EGS程序。