首页> 外文期刊>Review of Scientific Instruments >Calibration of a flat field soft x-ray grating spectrometer for laser produced plasmas a
【24h】

Calibration of a flat field soft x-ray grating spectrometer for laser produced plasmas a

机译:用于激光产生的等离子体 a 的平场软X射线光栅光谱仪的校准

获取原文
获取原文并翻译 | 示例
           

摘要

We have calibrated the x-ray response of a variable line spaced grating spectrometer, known as the VSG, at the Fusion and Astrophysics Data and Diagnostic Calibration Facility at the Lawrence Livermore National Laboratory (LLNL). The VSG has been developed to diagnose laser produced plasmas, such as those created at the Jupiter Laser Facility and the National Ignition Facility at LLNL and at both the Omega and Omega EP lasers at the University of Rochester’s Laboratory for Laser Energetics. The bandwidth of the VSG spans the range of ∼ 6–60 Å. The calibration results presented here include the VSG’s dispersion and quantum efficiency. The dispersion is determined by measuring the x rays emitted from the hydrogenlike and heliumlike ions of carbon, nitrogen, oxygen, neon, and aluminum. The quantum efficiency is calibrated to an accuracy of 30% or better by normalizing the x-ray intensities recorded by the VSG to those simultaneously recorded by an x-ray microcalorimeter spectrometer. © 2010 American Institute of Physics Article Outline INTRODUCTION CALIBRATION ARRANGEMENT RESULT
机译:我们已经在劳伦斯·利弗莫尔国家实验室(LLNL)的融合与天体数据和诊断校准设施处,对可变线距光栅光谱仪(称为VSG)的X射线响应进行了校准。 VSG的开发用于诊断激光产生的等离子体,例如在LLNL的Jupiter激光工厂和国家点火工厂以及在罗彻斯特大学激光能量实验室的Omega和Omega EP激光器中产生的等离子体。 VSG的带宽范围约为6–60Å。此处显示的校准结果包括VSG的色散和量子效率。通过测量从碳,氮,氧,氖和铝的氢和氦样离子发出的X射线来确定色散。通过将VSG记录的X射线强度归一化为X射线微热量计光谱仪同时记录的强度,可以将量子效率校准为30%或更高的精度。 ©2010美国物理研究所文章大纲简介校准校准结果

著录项

  • 来源
    《Review of Scientific Instruments》 |2010年第10期|p.1-3|共3页
  • 作者单位

    Lawrence Livermore National Laboratory, P.O. Box 808, Livermore, California 94551-0808, USA|University of California, Davis, One Shields Avenue, Davis, California 95616, USA;

    Lawrence Livermore National Laboratory, P.O. Box 808, Livermore, California 94551-0808, USA;

    Lawrence Livermore National Laboratory, P.O. Box 808, Livermore, California 94551-0808, USA;

    University of California, Davis, One Shields Avenue, Davis, California 95616, USA;

    Lawrence Livermore National Laboratory, P.O. Box 808, Livermore, California 94551-0808, USA;

    Lawrence Livermore National Laboratory, P.O. Box 808, Livermore, California 94551-0808, USA|University of California, Davis, One Shields Avenue, Davis, California 95616, USA;

    NASA Goddard Space Flight Center, Greenbelt, Maryland 20770, USA;

    NASA Goddard Space Flight Center, Greenbelt, Maryland 20770, USA;

    Lawrence Livermore National Laboratory, P.O. Box 808, Livermore, California 94551-0808, USA;

    Lawrence Livermore National Laboratory, P.O. Box 808, Livermore, California 94551-0808, USA;

    NASA Goddard Space Flight Center, Greenbelt, Maryland 20770, USA;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    calibration; plasma diagnostics; plasma temperature; plasma X-ray sources; X-ray spectrometers;