A Ratio‐Recording Double Beam Infra‐Red Spectrophotometer Using Phase Discrimination and a Single Detector

机译:使用相位鉴别和单个检测器的比率记录双光束红外分光光度计

获取原文
获取原文并翻译 | 示例

摘要

A simple modification is described for converting a Perkin‐Elmer Infra‐Red Spectrophotometer to double beam operation for direct recording of percent transmission versus wave‐length. A phase discrimination system is used, requiring no changes in the optical system, and no sacrifice of versatility or ease of prism interchange. A complete spectrum over the rocksalt region may be obtained in 25 minutes with good resolution, and with no attention from an operator. Samples as small as 1×6 mm can be studied. The system is readily adaptable to other spectrophotometers, including those in the visual and ultraviolet range, the only limitation being sufficiently rapid response by the detector.
机译:描述了一个简单的修改,将Perkin-Elmer红外分光光度计转换为双光束操作,可直接记录透射率与波长的百分比。使用了鉴相系统,不需要改变光学系​​统,也无需牺牲多功能性或棱镜更换的便利性。可以在25分钟内以良好的分辨率获得岩石盐区域的完整光谱,并且无需操作人员的注意。可以研究小至1×6 mm的样品。该系统很容易适应其他分光光度计,包括可见光和紫外线范围内的分光光度计,唯一的限制是检测器的响应速度足够快。